3-20
Troubleshooting
Execution Error
-232
Invalid format
Indicates that a legal program data element was parsed but
could not be executed because the data format or structure is
inappropriate. For example, when loading memory tables or
when sending a
SYSTem:SET
parameter for an unknown
instrument.
-233
Invalid version
Indicates that a legal program data element was parsed but
could not be executed because the version of the data is
incorrect to the device. This particular error is used when file or
block data elements are recognized by the instrument, but
cannot be executed for reasons of version incompatibility. For
example, a non- supported file version or a non- supported
instrument version.
-240
Hardware error
Indicates that a legal program command or query could not be
executed because of a hardware problem in the device. The
definition of what constitutes a hardware problem is completely
device-specific. This error is used when the device cannot
detect the more specific errors described for errors -241
through -249.
-241
Hardware missing
Indicates that a legal program command or query could not be
executed because of missing device hardware. For example, an
option was not installed.
-250
Mass storage error
Indicates that a mass storage error has occurred. This message
is used when a device cannot detect the more specific errors
described for errors -251 through -259.
-252
Missing media
Indicates that a legal program command or query could not be
executed because of missing media, for instance no disk in the
disk drive. The definition of what constitutes missing media is
device-specific.
Содержание ESG Series
Страница 3: ...ii ...
Страница 29: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Страница 43: ...2 14 Using Functions Saving and Recalling an Instrument State ...
Страница 95: ...3 52 Troubleshooting HP Sales and Service Offices ...
Страница 109: ...4 14 Front and Rear Panel Rear Panel Overview ...
Страница 111: ...5 2 Hardkey and Softkey Reference ...
Страница 135: ...5 26 Hardkey and Softkey Reference Ampl ...
Страница 137: ...5 28 Hardkey and Softkey Reference Amplitude ...
Страница 139: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Страница 141: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Страница 161: ...5 52 Hardkey and Softkey Reference FM ...
Страница 169: ...5 60 Hardkey and Softkey Reference Frequency ...
Страница 171: ...5 62 Hardkey and Softkey Reference Help ...
Страница 173: ...5 64 Hardkey and Softkey Reference Hold ...
Страница 175: ...5 66 Hardkey and Softkey Reference Incr Set ...
Страница 195: ...5 86 Hardkey and Softkey Reference LF Out ...
Страница 197: ...5 88 Hardkey and Softkey Reference Local ...
Страница 199: ...5 90 Hardkey and Softkey Reference Mod On Off ...
Страница 200: ...5 91 Hardkey and Softkey Reference Mode Mode This front panel key is reserved for future use ...
Страница 201: ...5 92 Hardkey and Softkey Reference Mode ...
Страница 203: ...5 94 Hardkey and Softkey Reference Numeric Keypad ...
Страница 225: ...5 116 Hardkey and Softkey Reference Power Switch ...
Страница 235: ...5 126 Hardkey and Softkey Reference Preset ...
Страница 241: ...5 132 Hardkey and Softkey Reference Pulse ...
Страница 245: ...5 136 Hardkey and Softkey Reference Recall ...
Страница 247: ...5 138 Hardkey and Softkey Reference Return ...
Страница 249: ...5 140 Hardkey and Softkey Reference RF On Off ...
Страница 255: ...5 146 Hardkey and Softkey Reference Save ...
Страница 275: ...5 166 Hardkey and Softkey Reference Sweep List ...
Страница 311: ...6 10 Options and Accessories Upgrade and Retrofit Kits ...
Страница 312: ...7 1 7 Operation This chapter is reserved for future use ...
Страница 313: ...7 2 Operation ...
Страница 322: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...
Страница 341: ...9 18 Specifications HP ESG Series Signal Generators Specifications ...