2-11
Using Functions
Creating a Step Sweep and a List Sweep
Turning On Continuous Step Sweep
1. Press
Return
to move up one menu level.
2. Press the
Sweep
softkey. Another menu is displayed showing you choices for sweeping
either the frequency, amplitude, or frequency and amplitude data. Press the
Freq&Ampl
softkey. Selecting this softkey returns you to the previous menu and turns the sweep
function on.
3. Press the
Sweep Repeat Single Cont
softkey to toggle from
Single
to
Cont
. Notice
that the
SWEEP
display annunciator is turned on indicating that the signal generator is
sweeping.
4. Press the front panel
RF On/Off
key to toggle RF on. Notice that the display annunciator
changes from
RF OFF
to
RF ON
. The swept RF signal is now available at the RF
OUTPUT connector.
Configuring a List Sweep Using Step Sweep Data
1. Press the
Sweep Type List Step
softkey to toggle from
Step
to
List
.
2. Press the
Configure List Sweep
softkey. Another menu is displayed with softkeys that
you will use to create the sweep points. Notice that the display shows the current list
data (when no list has been previously created, the default is one point, set to the signal
generator’s maximum frequency,
−
135 dBm, with a dwell time of 2 ms).
3. Press
More (1 of 2)
,
Load List From Step Sweep
,
Confirm Load From Step Sweep
.
The points you defined in the step sweep are automatically loaded into the list.
Editing List Sweep Points
1. Change the dwell time for point 1 to 100 ms. Press the right arrow key twice until the
dwell time is highlighted. Press
More (2 of 2)
,
Edit Item
. The dwell time for point 1
becomes the active function. Enter 100 ms using the numeric keypad and pressing the
msec
terminator softkey. Notice that the next item in the table (in this case the
frequency value for point 2) becomes highlighted after pressing the terminator softkey.
2. Change the frequency for point 4 to 560 MHz. Press the down arrow key two times until
the frequency is highlighted. Press the
Edit Item
softkey. The frequency for point 4
becomes the active function. Enter 560 MHz using the numeric keypad and pressing the
MHz
terminator softkey.
3. Add a new point between points 7 and 8. Press the down arrow until any column in the
point 8 row is highlighted. Press the
Insert Row
softkey. A copy of point 8 has been
placed between points 7 and 8, creating a new point 8, and renumbering the successive
points.
Содержание ESG Series
Страница 3: ...ii ...
Страница 29: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Страница 43: ...2 14 Using Functions Saving and Recalling an Instrument State ...
Страница 95: ...3 52 Troubleshooting HP Sales and Service Offices ...
Страница 109: ...4 14 Front and Rear Panel Rear Panel Overview ...
Страница 111: ...5 2 Hardkey and Softkey Reference ...
Страница 135: ...5 26 Hardkey and Softkey Reference Ampl ...
Страница 137: ...5 28 Hardkey and Softkey Reference Amplitude ...
Страница 139: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Страница 141: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Страница 161: ...5 52 Hardkey and Softkey Reference FM ...
Страница 169: ...5 60 Hardkey and Softkey Reference Frequency ...
Страница 171: ...5 62 Hardkey and Softkey Reference Help ...
Страница 173: ...5 64 Hardkey and Softkey Reference Hold ...
Страница 175: ...5 66 Hardkey and Softkey Reference Incr Set ...
Страница 195: ...5 86 Hardkey and Softkey Reference LF Out ...
Страница 197: ...5 88 Hardkey and Softkey Reference Local ...
Страница 199: ...5 90 Hardkey and Softkey Reference Mod On Off ...
Страница 200: ...5 91 Hardkey and Softkey Reference Mode Mode This front panel key is reserved for future use ...
Страница 201: ...5 92 Hardkey and Softkey Reference Mode ...
Страница 203: ...5 94 Hardkey and Softkey Reference Numeric Keypad ...
Страница 225: ...5 116 Hardkey and Softkey Reference Power Switch ...
Страница 235: ...5 126 Hardkey and Softkey Reference Preset ...
Страница 241: ...5 132 Hardkey and Softkey Reference Pulse ...
Страница 245: ...5 136 Hardkey and Softkey Reference Recall ...
Страница 247: ...5 138 Hardkey and Softkey Reference Return ...
Страница 249: ...5 140 Hardkey and Softkey Reference RF On Off ...
Страница 255: ...5 146 Hardkey and Softkey Reference Save ...
Страница 275: ...5 166 Hardkey and Softkey Reference Sweep List ...
Страница 311: ...6 10 Options and Accessories Upgrade and Retrofit Kits ...
Страница 312: ...7 1 7 Operation This chapter is reserved for future use ...
Страница 313: ...7 2 Operation ...
Страница 322: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...
Страница 341: ...9 18 Specifications HP ESG Series Signal Generators Specifications ...