Using Functions
Setting Up Internally-Generated Amplitude Modulation
Setting Up Internally-Generated Amplitude Modulation
Using this procedure, you will learn how to generate an amplitude-modulated
signal with the following characteristics:
•
carrier frequency set to 1340 kHz
•
power level set to 0 dBm
•
AM depth set to 90%
•
AM rate set to 10 kHz
Setting the Carrier Frequency
1. Press
Preset
to return the signal generator to the factory-defined instrument state.
2. Press the front panel
Frequency
key. Frequency becomes the active function and the
normal preset value for frequency is displayed in the active entry area.
3. Enter 1340 kHz using the numeric keypad and pressing the
kHz
terminator softkey. The
new carrier frequency is shown in the frequency area of the display. You should see
1.340 000 00 MHz
.
Setting the Power Level
1. Press the front panel
Amplitude
key. Amplitude becomes the active function and the
normal preset value for amplitude is displayed in the active entry area.
2. Enter 0 dBm using the numeric keypad and pressing the
dBm
terminator softkey. The
new power level is shown in the amplitude area of the display. You should see
0.00 dBm
.
Setting the AM Depth
1. Press the front panel
AM
key. The first level menu of softkeys is displayed.
2. Press the
AM Depth
softkey. AM depth becomes the active function and the normal
preset value for AM depth is displayed in the active entry area.
3. Enter 90% using the numeric keypad and pressing the
%
terminator softkey. The new
AM depth is displayed below the
AM Depth
softkey. You should see
90.0 %
in the
second line of the softkey.
Содержание ESG Series
Страница 3: ...ii ...
Страница 29: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Страница 43: ...2 14 Using Functions Saving and Recalling an Instrument State ...
Страница 95: ...3 52 Troubleshooting HP Sales and Service Offices ...
Страница 109: ...4 14 Front and Rear Panel Rear Panel Overview ...
Страница 111: ...5 2 Hardkey and Softkey Reference ...
Страница 135: ...5 26 Hardkey and Softkey Reference Ampl ...
Страница 137: ...5 28 Hardkey and Softkey Reference Amplitude ...
Страница 139: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Страница 141: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Страница 161: ...5 52 Hardkey and Softkey Reference FM ...
Страница 169: ...5 60 Hardkey and Softkey Reference Frequency ...
Страница 171: ...5 62 Hardkey and Softkey Reference Help ...
Страница 173: ...5 64 Hardkey and Softkey Reference Hold ...
Страница 175: ...5 66 Hardkey and Softkey Reference Incr Set ...
Страница 195: ...5 86 Hardkey and Softkey Reference LF Out ...
Страница 197: ...5 88 Hardkey and Softkey Reference Local ...
Страница 199: ...5 90 Hardkey and Softkey Reference Mod On Off ...
Страница 200: ...5 91 Hardkey and Softkey Reference Mode Mode This front panel key is reserved for future use ...
Страница 201: ...5 92 Hardkey and Softkey Reference Mode ...
Страница 203: ...5 94 Hardkey and Softkey Reference Numeric Keypad ...
Страница 225: ...5 116 Hardkey and Softkey Reference Power Switch ...
Страница 235: ...5 126 Hardkey and Softkey Reference Preset ...
Страница 241: ...5 132 Hardkey and Softkey Reference Pulse ...
Страница 245: ...5 136 Hardkey and Softkey Reference Recall ...
Страница 247: ...5 138 Hardkey and Softkey Reference Return ...
Страница 249: ...5 140 Hardkey and Softkey Reference RF On Off ...
Страница 255: ...5 146 Hardkey and Softkey Reference Save ...
Страница 275: ...5 166 Hardkey and Softkey Reference Sweep List ...
Страница 311: ...6 10 Options and Accessories Upgrade and Retrofit Kits ...
Страница 312: ...7 1 7 Operation This chapter is reserved for future use ...
Страница 313: ...7 2 Operation ...
Страница 322: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...
Страница 341: ...9 18 Specifications HP ESG Series Signal Generators Specifications ...