2-13
Using Functions
Saving and Recalling an Instrument State
Saving and Recalling an Instrument State
Using this procedure, you will learn how to save instrument settings to a memory
register and to recall the settings.
1. Press
Preset
to return the signal generator to the factory-defined instrument state.
2. Set up the signal generator with the following changes:
a. Set the frequency to 800 MHz.
b. Set the amplitude to 0 dBm.
c. Enable amplitude modulation (
AM
display annunciator is on).
3. Save this instrument state in the signal generator memory in sequence 1, memory
register 01. Press the front panel
Save
key and then press the
Select Seq
softkey. The
sequence number becomes the active function. The signal generator will display the last
sequence that you have used. Set the sequence to 1 using the arrow keys.
Press the
Select Reg
softkey. The register number in sequence 1 becomes the active
function. The signal generator will either display the last register used [accompanied
by the text:
(in use)
] or, if no registers are in use, will display register 00
[accompanied by the text:
(available)
]. Use the arrow keys to select register 01.
4. Press the
Save Seq[1] Reg[01]
softkey. The current instrument settings including the
frequency, amplitude, and modulation changes you made have been stored in signal
generator memory.
5. Press
Preset
to return the signal generator to the factory-defined instrument state.
6. Recall your instrument state. Press the front panel
Recall
key. Notice that the
Select Seq
softkey shows sequence 1. (This is the last sequence that you have used.)
You do not need to change the sequence. Press
RECALL Reg
. The register to be
recalled in sequence 1 becomes the active function. Press the up arrow key once to
select register 1. Notice that your stored instrument settings have been immediately
recalled.
Содержание ESG Series
Страница 3: ...ii ...
Страница 29: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Страница 43: ...2 14 Using Functions Saving and Recalling an Instrument State ...
Страница 95: ...3 52 Troubleshooting HP Sales and Service Offices ...
Страница 109: ...4 14 Front and Rear Panel Rear Panel Overview ...
Страница 111: ...5 2 Hardkey and Softkey Reference ...
Страница 135: ...5 26 Hardkey and Softkey Reference Ampl ...
Страница 137: ...5 28 Hardkey and Softkey Reference Amplitude ...
Страница 139: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Страница 141: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Страница 161: ...5 52 Hardkey and Softkey Reference FM ...
Страница 169: ...5 60 Hardkey and Softkey Reference Frequency ...
Страница 171: ...5 62 Hardkey and Softkey Reference Help ...
Страница 173: ...5 64 Hardkey and Softkey Reference Hold ...
Страница 175: ...5 66 Hardkey and Softkey Reference Incr Set ...
Страница 195: ...5 86 Hardkey and Softkey Reference LF Out ...
Страница 197: ...5 88 Hardkey and Softkey Reference Local ...
Страница 199: ...5 90 Hardkey and Softkey Reference Mod On Off ...
Страница 200: ...5 91 Hardkey and Softkey Reference Mode Mode This front panel key is reserved for future use ...
Страница 201: ...5 92 Hardkey and Softkey Reference Mode ...
Страница 203: ...5 94 Hardkey and Softkey Reference Numeric Keypad ...
Страница 225: ...5 116 Hardkey and Softkey Reference Power Switch ...
Страница 235: ...5 126 Hardkey and Softkey Reference Preset ...
Страница 241: ...5 132 Hardkey and Softkey Reference Pulse ...
Страница 245: ...5 136 Hardkey and Softkey Reference Recall ...
Страница 247: ...5 138 Hardkey and Softkey Reference Return ...
Страница 249: ...5 140 Hardkey and Softkey Reference RF On Off ...
Страница 255: ...5 146 Hardkey and Softkey Reference Save ...
Страница 275: ...5 166 Hardkey and Softkey Reference Sweep List ...
Страница 311: ...6 10 Options and Accessories Upgrade and Retrofit Kits ...
Страница 312: ...7 1 7 Operation This chapter is reserved for future use ...
Страница 313: ...7 2 Operation ...
Страница 322: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...
Страница 341: ...9 18 Specifications HP ESG Series Signal Generators Specifications ...