5-149
Hardkey and Softkey Reference
Sweep/List
Ampl Stop
Press this softkey to set the amplitude of the last point in the sweep. You can set
the value anywhere in the range of the signal generator’s specified output power.
Press
Ampl Stop
and amplitude becomes the active function for the last point in
the sweep. To enter the value, rotate the front panel knob until the desired value is
displayed, use the up and down arrow keys, or enter the value using the numeric
keypad and press the
dBm
,
dBuV
,
dBuVemf
,
mV
,
uV
,
mVemf
, or
uVemf
terminator
softkey.
Softkey Location: Press
Sweep/List
,
Configure Step Sweep
,
Ampl Stop
Status after Normal Preset:
−
135 dBm
SCPI Commands:
[:SOURce]:POWer:STOP <val><unit>
[:SOURce]:POWer:STOP?
Bus
This softkey is one of the choices in both the Sweep Trigger menu and the Point
Trigger menu. Press either the
Sweep Trigger
softkey or the
Point Trigger
softkey
to view these menus. Choosing
Bus
in the Sweep Trigger menu allows you to
trigger a list or step sweep using the HP-IB. Choosing
Bus
in the Point Trigger
menu allows you to trigger a sweep point-by-point using the HP-IB via *TRG or
the GET line (Group Execute Trigger).
Softkey Location: Press
Sweep/List
,
More (1 of 2)
,
Sweep Trigger
,
Bus
or press
Sweep/List
,
More (1 of 2)
,
Point Trigger
,
Bus
Status after Normal Preset: Sweep and point trigger are set to Immediate
SCPI Commands:
[:SOURce]:LIST:TRIGger:SOURce BUS
[:SOURce]:LIST:TRIGger:SOURce?
Содержание ESG Series
Страница 3: ...ii ...
Страница 29: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Страница 43: ...2 14 Using Functions Saving and Recalling an Instrument State ...
Страница 95: ...3 52 Troubleshooting HP Sales and Service Offices ...
Страница 109: ...4 14 Front and Rear Panel Rear Panel Overview ...
Страница 111: ...5 2 Hardkey and Softkey Reference ...
Страница 135: ...5 26 Hardkey and Softkey Reference Ampl ...
Страница 137: ...5 28 Hardkey and Softkey Reference Amplitude ...
Страница 139: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Страница 141: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Страница 161: ...5 52 Hardkey and Softkey Reference FM ...
Страница 169: ...5 60 Hardkey and Softkey Reference Frequency ...
Страница 171: ...5 62 Hardkey and Softkey Reference Help ...
Страница 173: ...5 64 Hardkey and Softkey Reference Hold ...
Страница 175: ...5 66 Hardkey and Softkey Reference Incr Set ...
Страница 195: ...5 86 Hardkey and Softkey Reference LF Out ...
Страница 197: ...5 88 Hardkey and Softkey Reference Local ...
Страница 199: ...5 90 Hardkey and Softkey Reference Mod On Off ...
Страница 200: ...5 91 Hardkey and Softkey Reference Mode Mode This front panel key is reserved for future use ...
Страница 201: ...5 92 Hardkey and Softkey Reference Mode ...
Страница 203: ...5 94 Hardkey and Softkey Reference Numeric Keypad ...
Страница 225: ...5 116 Hardkey and Softkey Reference Power Switch ...
Страница 235: ...5 126 Hardkey and Softkey Reference Preset ...
Страница 241: ...5 132 Hardkey and Softkey Reference Pulse ...
Страница 245: ...5 136 Hardkey and Softkey Reference Recall ...
Страница 247: ...5 138 Hardkey and Softkey Reference Return ...
Страница 249: ...5 140 Hardkey and Softkey Reference RF On Off ...
Страница 255: ...5 146 Hardkey and Softkey Reference Save ...
Страница 275: ...5 166 Hardkey and Softkey Reference Sweep List ...
Страница 311: ...6 10 Options and Accessories Upgrade and Retrofit Kits ...
Страница 312: ...7 1 7 Operation This chapter is reserved for future use ...
Страница 313: ...7 2 Operation ...
Страница 322: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...
Страница 341: ...9 18 Specifications HP ESG Series Signal Generators Specifications ...