2-12
Using Functions
Creating a Step Sweep and a List Sweep
4. Create a new point between points 10 and 11. Use the arrow keys to highlight the
frequency for point 11. Press the
Insert Item
softkey. Notice that a new frequency item
is placed at point 11 and the frequency item previously occupying that position has
shifted down to point 12. The power and dwell time items do not shift down. Press the
Edit Item
softkey. Enter 700 MHz using the numeric keypad and pressing the
MHz
terminator softkey.
NOTE:
An informational
ERR
annunciator is turned on at this time indicating that the frequency
and power lists are of unequal size. You will correct that problem in the following steps.
The annunciator will not turn off until you clear the error queue in the
Utility
menu.
The power for point 11 should now be highlighted. Press the
Insert Item
softkey, then
press the
Edit Item
softkey and enter 1 dBm using the numeric keypad and pressing
the
dBm
terminator softkey. A new power item is placed at point 11 and the power
item previously occupying that position has shifted down to point 12.
The dwell time for point 11 should now be highlighted. Press the
Insert Item
softkey.
The dwell time from the previous point 11 has been copied into the new point 11. A
new dwell time item is placed at point 11 and the dwell time item previously
occupying that position has shifted down to point 12.
Turning On List Sweep for a Single Sweep
1. Press
Return
to move up one menu level.
2. Notice that the
Sweep
softkey is still set to sweep both frequency and amplitude data.
You do not need to change it.
3. Press the
Sweep Repeat Single Cont
softkey to toggle from
Cont
to
Single
. Notice
that the
SWEEP
display annunciator is turned off. The sweep will not occur until it is
triggered.
4. Press
Single Sweep
. The signal generator will sweep the points in your list once.
Notice that the
SWEEP
display annunciator is turned on during the sweep.
5. Change the sweep trigger to occur when you press the front panel
Trigger
key. Press
More (1 of 2)
,
Sweep Trigger
. Another menu is displayed showing you choices for
triggering a sweep. Press the
Trigger Key
softkey.
6. Press
More (2 of 2)
,
Single Sweep
to arm the sweep. Notice that the
ARMED
display
annunciator has turned on.
7. Press the front panel
Trigger
key. The signal generator will sweep the points in your list
once and the
SWEEP
display annunciator is turned on during the sweep.
Содержание ESG Series
Страница 3: ...ii ...
Страница 29: ...1 12 Preparing for Use Verifying Signal Generator Operation ...
Страница 43: ...2 14 Using Functions Saving and Recalling an Instrument State ...
Страница 95: ...3 52 Troubleshooting HP Sales and Service Offices ...
Страница 109: ...4 14 Front and Rear Panel Rear Panel Overview ...
Страница 111: ...5 2 Hardkey and Softkey Reference ...
Страница 135: ...5 26 Hardkey and Softkey Reference Ampl ...
Страница 137: ...5 28 Hardkey and Softkey Reference Amplitude ...
Страница 139: ...5 30 Hardkey and Softkey Reference Arrow Keys ...
Страница 141: ...5 32 Hardkey and Softkey Reference Display Contrast Keys ...
Страница 161: ...5 52 Hardkey and Softkey Reference FM ...
Страница 169: ...5 60 Hardkey and Softkey Reference Frequency ...
Страница 171: ...5 62 Hardkey and Softkey Reference Help ...
Страница 173: ...5 64 Hardkey and Softkey Reference Hold ...
Страница 175: ...5 66 Hardkey and Softkey Reference Incr Set ...
Страница 195: ...5 86 Hardkey and Softkey Reference LF Out ...
Страница 197: ...5 88 Hardkey and Softkey Reference Local ...
Страница 199: ...5 90 Hardkey and Softkey Reference Mod On Off ...
Страница 200: ...5 91 Hardkey and Softkey Reference Mode Mode This front panel key is reserved for future use ...
Страница 201: ...5 92 Hardkey and Softkey Reference Mode ...
Страница 203: ...5 94 Hardkey and Softkey Reference Numeric Keypad ...
Страница 225: ...5 116 Hardkey and Softkey Reference Power Switch ...
Страница 235: ...5 126 Hardkey and Softkey Reference Preset ...
Страница 241: ...5 132 Hardkey and Softkey Reference Pulse ...
Страница 245: ...5 136 Hardkey and Softkey Reference Recall ...
Страница 247: ...5 138 Hardkey and Softkey Reference Return ...
Страница 249: ...5 140 Hardkey and Softkey Reference RF On Off ...
Страница 255: ...5 146 Hardkey and Softkey Reference Save ...
Страница 275: ...5 166 Hardkey and Softkey Reference Sweep List ...
Страница 311: ...6 10 Options and Accessories Upgrade and Retrofit Kits ...
Страница 312: ...7 1 7 Operation This chapter is reserved for future use ...
Страница 313: ...7 2 Operation ...
Страница 322: ...8 9 Safety and Regulatory Declaration of Conformity Declaration of Conformity ...