Adjustment Procedure
1. Reset the AFG:
*RST;*CLS
2. Set up the AFG to output a 1 V peak, 100 kHz sinewave:
VOLT 1VPK;
:FREQ 100E3
INIT:IMM
3. Set the power meter expected frequency to 100 kHz and measure the
AFG output. Convert the reading to RMS volts and note the result for
use in step 5.
4. Connect the DMM to the AFG Output as shown in Figure 3-3:
5. Measure the AFG Output with the DMM. Calculate the correction
factor as shown below and note the result for use in step 10:
C.F.
=
DMM reading
Power
Meter
reading
(
Step
3
)
6. Start the calibration procedure:
CAL:AC:BEGIN
Initiate AC cal
Figure 3-3. Flatness Adjustment Setup (DMM)