Test Procedure
1. Reset the AFG:
*RST;*CLS
Reset AFG and clear status registers
Perform steps 2 - 6 for each frequency listed in Table 2-5:
2. Set the AFG to output a sinewave
VOLT
<amplitude>
DBM
where <amplitude> is the value specified in Table 2-5.
3. Set the AFG output frequency:
ABORT
FREQ
<frequency>
Set frequency
INIT:IMM
Initiate waveform
where <frequency> is the value specified in Table 2-5.
4. Set the spectrum analyzer center frequency to the output frequency of
the AFG. Measure the peak amplitude of the fundamental in dBm.
Record the reading in Table 2-5.
5. Set the spectrum analyzer center frequency to the second harmonic
( 2
×
fundamental frequency ). Measure the peak amplitude of the
second harmonic. Repeat for third through fifth harmonics. Record
the amplitude of each harmonic in Table 2-5.
6. Calculate total harmonic distortion as shown below. Record the
results in Table 2-5. Record the final result in Table 2-7.
thd
(
dBc
)
=
20
⋅
log
√
result
2
2
+
result
3
2
+
result
4
2
+
result
5
2
where
result
n
(
volts
)
=
10
[
(
n
th
Harmonic
(
dBm
)
−
Fundamental
(
dBm
)
)
/20
]
50 Verification Tests
Chapter 2