Test 2-6: Total Harmonic Distortion
Description
The purpose of this test is to verify that the AFG meets its specifications for
sine wave total harmonic distortion (THD).
Equipment Setup
–
Connect equipment as shown in Figure 2-9
–
Set Spectrum Analyzer to:
Ref Level = 25 dBm
Freq Span = 1 kHz
Resolution BW = 30 Hz
Video BW = 30 Hz
NOTE
These are recommended settings only. Adjust your spectrum analyzer as
necessary.
Figure 2-9. Total Harmonic Distortion Test Setup
Chapter 2
Verification Tests 49