Repeat steps 3 - 5 until the AFG returns 0 (zero) for the current
calibration point:
NOTE
The AFG will cycle through calibration points 1 through 14 until the
readings are within range. The AFG signals that the calibration has been
successfully completed by returning a 0 (zero) for the current calibration
point.
3. If the current calibration point is 6, set the DMM range to 10 V to
prevent an overload when the AFG sets itself for the next reading.
This should be done immediately after performing the DMM
measurement for cal point 6 (and before sending the reading to the
AFG).
4. Trigger the DMM and note the reading.
5. Send the reading to the AFG:
CAL:DC:POINT? <
reading
>
where <reading> is the DMM reading from step 4. The AFG will return, in
order, the number of the current calibration point and an error code. Any
non-zero error code indicates a failure.
NOTE
If an error occurs, start over from the beginning. If the error persists, factory
calibration may be required.
66 Adjustments
Chapter
3