9. Set the power meter expected frequency to the crossover frequency
(100 kHz). Measure the AFG output power and convert the reading
to volts. Record the reading in Table 2-4.
Reading
(
volts
)
=(
√
Reading
(
watts
)
×
50
)
10. Calculate the correction factor that will be used to reference the
power meter to the DMM. Record the correction factor in Table 2-4.
Correction Factor
=
DMM reading
at
100
kHz
(
step
6
)
Power Meter reading
at
100
kHz
(
step 9
)
Repeat 11 - 14 for each frequency in Table 2-4:
11. Set the AFG output to the frequency specified in Table 2-4:
ABORT
FREQ
<frequency>
Set frequency
INIT:IMM
Initiate waveform
where <frequency> is the value specified in Table 2-4.
12. Set the power meter expected frequency to the AFG output frequency.
13. Measure the output power with the power meter, convert the reading
to volts, and multiply by the correction factor. Record the results in
Table 2-4.
Reading
(
volts
)
=(
√
Reading
(
watts
)
×
50
)
Corrected Reading
(
volts
)
=
Reading
(
volts
)
×
C.F.
(
step 10
)
14. Convert the reading to dBm. Calculate and record the error relative
to the reference level calculated in step 4. Record the results in
Table 2-4. Record the final result in Table 2-7.
Reading
(
dBm
)
=
20
×
log
Corrected Reading
(
volts
)
+
13.0103
Error
(
dB
)
=
Reading
(
dBm
)
−
Reference
Level
(
dBm
)
Chapter 2
Verification Tests 41
Содержание E1340A
Страница 6: ...Notes 6 HP E1340A Arbitrary Function Generator Service Manual ...
Страница 8: ......
Страница 10: ...Notes 12 What s in this Manual HP E1340A Service Manual ...
Страница 18: ...Notes 20 General Information Chapter 1 ...
Страница 74: ...Notes 76 Adjustments Chapter 3 ...
Страница 78: ...Notes 80 Replaceable Parts Chapter 4 ...
Страница 104: ...Notes 106 Verification Tests C Programs Appendix A ...