Table 2-7. Performance Test Record for the HP E1340A (Page 5 of 5)
Model __________________________ Report No. _______________________ Date _______________
Test
Description
Minimum
Measured
Reading
Maximum
Meas
Uncert
TAR
Test 2-6. Total Harmonic Distortion Test (Values in dBc) *
Amplitude at 23 dBm:
100 kHz
1 MHz
10 MHz
Amplitude at 10 dBm:
100 kHz
1 MHz
10 MHz
____________
____________
____________
____________
____________
____________
-60 dBc
-49 dBc
-33 dBc
-63 dBc
-54 dBc
-47 dBc
1.7 dB
1.7 dB
1.7 dB
1.7 dB
1.7 dB
1.7 dB
NA
NA
NA
NA
NA
NA
Test 2-7. Spurious/Non-harmonic Distortion Test (Values in dBc) *
10 MHz, 10 dBm sinewave:
1 MHz - 9.5 MHz
10.5 MHz - 19 MHz
21 MHz - 29 MHz
31 MHz - 39 MHz
41 MHz - 49 MHz
51 MHz - 75 MHz
75 MHz - 100 MHz
100 MHz - 125 MHz
125 MHz - 150 MHz
____________
____________
____________
____________
____________
____________
____________
____________
____________
-43 dBc
-43 dBc
-43 dBc
-43 dBc
-43 dBc
-43 dBc
-43 dBc
-43 dBc
-43 dBc
1.23 dB
1.23 dB
1.23 dB
1.23 dB
1.23 dB
1.23 dB
1.23 dB
1.23 dB
1.23 dB
NA
NA
NA
NA
NA
NA
NA
NA
NA
* Single-sided test -- Minimum is not applicable
62 Verification Tests
Chapter 2
Содержание E1340A
Страница 6: ...Notes 6 HP E1340A Arbitrary Function Generator Service Manual ...
Страница 8: ......
Страница 10: ...Notes 12 What s in this Manual HP E1340A Service Manual ...
Страница 18: ...Notes 20 General Information Chapter 1 ...
Страница 74: ...Notes 76 Adjustments Chapter 3 ...
Страница 78: ...Notes 80 Replaceable Parts Chapter 4 ...
Страница 104: ...Notes 106 Verification Tests C Programs Appendix A ...