Test 2-7: Spurious/Non-Harmonic Distortion
Description
The purpose of this test is to verify that the AFG meets its specifications for
non-harmonic and spurious distortion.
Equipment Setup
–
Connect equipment as shown in Figure 2-10
–
Set Spectrum Analyzer to :
Ref Level = 10 dBm
Resolution BW = 3 kHz
Video BW = 3 kHz
NOTE
These are recommended settings only. Adjust your spectrum analyzer as
necessary.
Test Procedure
1. Reset the AFG:
*RST;*CLS
Reset AFG and clear status registers
Figure 2-10. Spurious/Non-Harmonic Distortion Test Setup
54 Verification Tests
Chapter 2