Test 2-3: AC Accuracy
Description
The purpose of this test is to verify that the AFG meets its specifications for
AC accuracy at 1 kHz.
Equipment Setup
–
Connect equipment as shown in Figure 2-5
–
Set DMM to ACV, autorange
Test Procedure
1. Reset the AFG:
*RST;*CLS
Reset AFG and clear status registers
2. Set up the AFG to output a 1 kHz sinewave:
FREQ 1E3;
Set freq to 1 kHz
:OUTP:LOAD INF
Infinite load
VOLT:UNIT VRMS
Set units to rms volts
INIT:IMM
Initiate waveform
Figure 2-5. AC Accuracy Test Setup
36 Verification Tests
Chapter 2