Performance Test Record
Table 2-7, "Performance Test Record for the HP E1340A AFG," is a form
you may copy and use to record performance verification test results for the
AFG. Table 2-7 shows AFG accuracy, measurement uncertainty, and test
accuracy ratio (TAR) values.
AFG Test Limits
Test limits are defined using the specifications in Appendix A of the
HP E1340A User’s Manual. The specifications for Total Harmonic
Distortion and Spurious/Non-harmonic Distortion are single-sided
(i.e., there is an upper limit but no lower limit). In the Performance
Test Record, the "Minimum" column will be blank.
Measurement
Uncertainty
For the performance verification tests in this manual, the measurement
uncertainties are based on the accuracy specifications for the following test
equipment:
Performance Test
Test Equipment
1. DC Accuracy
HP 3458A
2. DC Offset
HP 3458A
3. AC Accuracy
HP 3458A
4. AC Flatness*
HP 3458A
HP 8902A
5. Frequency Accuracy
HP 5334B
6. Total Harmonic Distortion
HP 3585B
7. Spurious/Non-harmonic Distortion
HP 8566B
* Includes following uncertainties: HP 8902A Range linearity, HP 11722A
Power Sensor Cal Factor uncertainty, HP 3458A accuracy at 100 kHz.
Test Accuracy
Ratio (TAR)
Test Accuracy Ratio (TAR) for the HP E1340A is defined as: AFG
Accuracy/Measurement Uncertainty, i.e.,
TAR
=
Maximum
−
Expected
Reading
Measurement
Uncertainty
For single-sided measurements, Test Accuracy Ratio is not defined, so
"NA" (Not Applicable) will appear in the TAR column. For TARs that
exceed 10:1, the entry is ">10:1".
Chapter 2
Verification Tests 57
Содержание E1340A
Страница 6: ...Notes 6 HP E1340A Arbitrary Function Generator Service Manual ...
Страница 8: ......
Страница 10: ...Notes 12 What s in this Manual HP E1340A Service Manual ...
Страница 18: ...Notes 20 General Information Chapter 1 ...
Страница 74: ...Notes 76 Adjustments Chapter 3 ...
Страница 78: ...Notes 80 Replaceable Parts Chapter 4 ...
Страница 104: ...Notes 106 Verification Tests C Programs Appendix A ...