380
OUTPUT 70903;"INIT"
390
!
Retrieve the readings from multimeter memory and enter them into the
400
!
controller. Clear the switchbox to exit the continuous scanning mode.
410
OUTPUT 70903;"FETC?"
420
ENTER 70903;Rdgs(*)
430
CLEAR 70914
440
END
Comments
•
The multimeter at secondary address 03 (logical address 24) is
connected to the switchbox at secondary address 14 (logical
address 112) with an analog bus cable and a digital bus cable.
•
The analog bus carries the input signals to the multimeter. The
digital bus is used to carry a "multimeter complete" signal to the
switchbox to trigger the next channel closing.
•
Because of the fast rate at which the FET channels close, the
multimeter is triggered once (
INIT
) and then samples continuously
(
SAMP:COUN 500
). Thus, the multimeter ignores the multiplexer
"channel closed" signal on the digital bus.
•
Using this configuration, the multimeter is able to continuously scan the
switchbox and store readings in its memory at a 76
µ
s (13 kHz) rate.
Chapter 3
Using the HP E1326B/E1411B Multimeter 51