:RESolution <resolution>
:RESolution? [MIN | MAX]
Note
The root command
[SENSe:]
is an implied command and can be omitted.
FUNCtion
[SENSe:]FUNCtion[:<
function
>]
selects the measurement function. You can
select 4-wire resistance, AC voltage, or DC voltage.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Units
<
function
>
discrete
:FRESistance
|
:VOLTage:AC |
:VOLTage[:DC]
none
Comments
•
The
:DC
parameter is optional. Both of the following command statements select
the DC voltage function:
FUNC:VOLT:DC
or
FUNC:VOLT
•
*RST Condition:
SENS:FUNC:VOLT:DC
Example
Changing Measurement Function
CONF:VOLT:DC
!Function: DC voltage; stand-alone
multimeter.
FUNC:FRES
!Change function to 4-wire resistance.
READ?
!Place multimeter in wait-for-trigger state and
make measurement; send reading to output
buffer.
enter statement
!Enter reading into computer.
FUNCtion?
[SENSe:]FUNCtion?
returns one of the following quoted strings to the output
buffer: “
FRES
”, “
VOLT:AC
”, or “
VOLT
”.
Example
Querying the Measurement Function
FUNC:FRES
!Function: 4-wire ohms; stand-alone
multimeter.
FUNC?
!Query multimeter to return selected
function.
enter statement
!Enter quoted string into computer.
Chapter 5
Command Reference 162