DISPlay
The
DISPlay
command subsystem monitors the state of the selected multiplexer
channel within the scanning multimeter. This command is useful only with
mainframes that have a front panel display, such as the HP 75000 Series B
Mainframe (Model HP E1301A).
Subsystem Syntax
DISPlay
:MONitor
:CHANnel <channel>
:CHANnel?
[:STATe] <mode>
[:STATe]?
:MONitor:CHANnel
DISPlay:MONitor:CHANnel <
channel
>
selects a single multiplexer channel to be
monitored. Use the
DISPlay:MONitor:STATe
command to enable and disable the
monitor mode.
Parameters
Parameter Name
Parameter Type
Range of Values
Default Units
<
channel
>
numeric
Refer to the multiplexer user’s manual for
a list of channels available.
none
Comments
•
Use the channel parameter to specify a single multiplexer channel within the
scanning multimeter. The channel list is of the form
(@ccnn)
, where
cc
= card
number and
nn
= channel number
(105 is channel 05 of card number 1, for example).
•
Use
AUTO
in place of channel to display measurements from the most recent
channel to receive a
CONFigure
or
MEASure
command. The channel number and
measurement are updated as the scan progresses. You may want to add a small
delay to the scan or use a slow, externally paced scan in order to view each
channel from the mainframe’s front panel.
•
The following example shows the multimeter’s monitor mode display on the HP
E1301A front panel.
•
Related Commands:
DISPlay:MONitor:STATe
•
*RST Condition:
DISP:MON:CHAN AUTO
VOLTMTR_8:
Chan:
101
+1.465302E-01
mon
volt
Chapter 5
Command Reference 136