Accuracy conditions for table below: Autozero on,, one hour warmup. Within 90 days
and
±
5
°
C of calibration temperature (module calibrated at 18
°
C to 28
°
C).
90-Day Accuracy vs. Aperture
±
(% of r ohms)
Range
267/320 ms
16.7/20 ms
2.5 ms
100
µ
s
10
µ
s
256
Ω
.035% + 10 m
Ω
.035% + 10 m
Ω
.035% + 10 m
Ω
.07% + 50 m
Ω
.12% + 50 m
Ω
2 k
Ω
.025% + 20 m
Ω
.025% + 20 m
Ω
.025% + 20 m
Ω
.06% + 150 m
Ω
.10% + 200 m
Ω
16 k
Ω
.025% + 200 m
Ω
.025% + 200 m
Ω
.025% + 200 m
Ω
.06% + 1
Ω
.1% + 2
Ω
131 k
Ω
.025% + 1
Ω
.025% + 1
Ω
.025% + 1
Ω
.06% + 8
Ω
.1% + 16
Ω
1 M
Ω
.025% + 10
Ω
.025% + 10
Ω
.025% + 10
Ω
.06% + 60
Ω
.1% + 120
Ω
Accuracy conditions for table below: Autozero on,, one hour warmup. Within 1 year
and
±
5
°
C of calibration temperature (module calibrated at 18
°
C to 28
°
C).
1-Year Accuracy vs. Aperture
±
(% of r ohms)
Range
267/320 ms
16.7/20 ms
2.5 ms
100
µ
s
10
µ
s
256
Ω
.05% + 10 m
Ω
.05% + 10 m
Ω
.05% + 10 m
Ω
.085% + 50 m
Ω
.135% + 50 m
Ω
2 k
Ω
.04% + 20 m
Ω
.04% + 20 m
Ω
.04% + 20 m
Ω
.075% + 150 m
Ω
.115% + 200 m
Ω
16 k
Ω
.04% + 200 m
Ω
.04% + 200 m
Ω
.04% + 200 m
Ω
.075% + 1
Ω
.115% + 2
Ω
131 k
Ω
.04% + 1
Ω
.04% + 1
Ω
.04% + 1
Ω
.075% + 8
Ω
.115% + 16
Ω
1 M
Ω
.04% + 10
Ω
.04% + 10
Ω
.04% + 10
Ω
.075% + 60
Ω
.115% + 120
Ω
Temperature Coefficient
±
(% of reading)/
°
C
Autozero Off Offset Error
(ohms)
Range
Temperature
Coefficient
10
µ
s
Aperture
Range
Additional
Offset Error
256
Ω
0.004
0.014
256
Ω
20 m
Ω
2 k
Ω
0.002
0.01
2 k
Ω
20 m
Ω
16 k
Ω
0.002
0.01
16 k
Ω
160 m
Ω
131 k
Ω
0.002
0.01
131 k
Ω
160 m
Ω
1 M
Ω
0.002
0.01
1 M
Ω
1.25
Ω
Conditions: 0
°
C to (cal temp -5
°
C)
,
(cal temp +5
°
C) to 55
°
C.
Conditions: Stable environment,
24 hours,
±
1
°
C.
Appendix A
HP E1326B/E1411B Multimeter Specifications 193
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com