2-wire thermistor (type = 2252, 5000, 10000).
OUTPUT 70903;"MEAS:TEMP? THER,type,(@channel_list)"
!
4-wire thermistor (type = 2252, 5000, 10000)
!
Channels available are 00 through 07.
OUTPUT 70903;"MEAS:TEMP? FTH,type,(@channel_list)"
!
2-wire RTD (type = 85, 92).
OUTPUT 70903;"MEAS:TEMP? RTD,type,(@channel_list)"
!
4-wire RTD (type = 85, 92)
!
Channels available are 00 through 07.
OUTPUT 70903;"MEAS:TEMP? FRTD,type,(@channel_list)"
The following statements can be substituted into the programs where the
multimeter configuration is set by
CONFigure
(and low-level commands).
!
AC voltage.
OUTPUT 70903;"CONF:VOLT:AC (@channel_list)"
!
2-wire resistance.
OUTPUT 70903;"CONF:RES (@channel_list)"
!
4-wire resistance (channels available are 00 through 07).
OUTPUT 70903;"CONF:FRES (@channel_list)"
!
Thermocouple (type = B, E, J, K, N14, N28, R, S, T).
OUTPUT 70903;"CONF:TEMP TC,type,(@channel_list)"
!
2-wire thermistor (type = 2252, 5000, 10000).
OUTPUT 70903;"CONF:TEMP THER,type,(@channel_list)"
!
4-wire thermistor (type = 2252, 5000, 10000)
!
Channels available are 00 through 07.
OUTPUT 70903;"CONF:TEMP FTH,type,(@channel_list)"
!
2-wire RTD (type = 85, 92).
OUTPUT 70903;"CONF:TEMP RTD,type,(@channel_list)"
!
4-wire RTD (type = 85, 92).
!
Channels available are 00 through 07.
OUTPUT 70903;"CONF:TEMP FRTD,type,(@channel_list)"
Note
The HP E1326B/E1411B multimeter also makes strain gage measurements.
Refer to the HP E1355A - E1358A Strain Gage Multiplexers User’s Manual
for example programs.
70 Using the HP E1326B/E1411B Multimeter
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com