parameter). This prevents the
MIN
or
MAX
resolution from being interpreted as a
range setting.
•
The fastest aperture time available when autoranging is 100
µ
s. In order to specify an
aperture time of 10
µ
s, you must select a fixed range.
•
Related Commands:
FETCh?
,
INITiate
,
READ?
Example
Making DC Voltage Measurements
CONF:VOLT 0.825,MAX,(@100:103)
!Function: DC voltage;
range selected: 0.91 V;
MAX resolution: 61.035
µ
V; specify
channel list.
TRIG:COUN 3
!Scan channel list 3 times
(take 4 readings per trigger).
INIT
!Place multimeter in wait-for-trigger state;
store readings in mainframe memory; trigger
source is IMMediate by default.
FETC?
!Place readings in output buffer.
enter statement
!Enter readings into computer.
133 Command Reference
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com