Accuracy conditions for table below: Autozero on, one hour warmup. Within 1 year
and
±
5
°
C of calibration temperature(module calibrated at 18
°
C to 28
°
C).
1-Year Accuracy vs. Aperture
±
(% of r volts)
Range
267/320 ms
16.7/20 ms
2.5 ms
100
µ
s
10
µ
s
125 mV
.033% + 5.0
µ
V
.033% + 5.0
µ
V .033% + 10
µ
V
.075% + 30
µ
V
.125% + 60
µ
V
1 V
.023% + 10
µ
V
.023% + 15
µ
V
.023% + 15
µ
V
.065% + 100
µ
V .110% + 200
µ
V
8 V
.020% + 50
µ
V
.020% + 50
µ
V
.020% + 80
µ
V
.065% + 750
µ
V .110% + 1.5 mV
64 V
.025% + 1.0 mV .025% + 1 mV
.025% + 1.0 mV .065% + 5.0 mV .110% + 20 mV
300 V
.025% + 5.0 mV .025% + 5 mV
.025% + 5.0 mV .065% + 30 mV
.110% + 80 mV
Temperature Coefficient
±
(% of reading)/
°
C
Autozero Off Offset Error
(volts)
Range
Temperature
Coefficient
10
µ
s
Aperture
Range
Additional
Offset Error
125 mV
0.003
0.013
125 mV
10
µ
V
1 V
0.001
0.01
1 V
10
µ
V
8 V
0.001
0.01
8 V
10
µ
V
64 V
0.001
0.01
64 V
1 mV
300 V
0.001
0.01
300 V
1 mV
Conditions: 0
°
C to (cal temp -5
°
C),
(cal temp +5
°
C) to 55
°
C.
Conditions: Stable environment,
24 hours,
±
1
°
C.
Appendix A
HP E1326B/E1411B Multimeter Specifications 191