Triggering Measurements
Triggering Measurements
The measurement cycle is started (triggered) by the occurrence of a trigger event.
The reliability and accuracy of the measurement result, as well as the speed of the
measurement cycle are influenced by the trigger mode in effect at the time the
trigger event occurs. Some modes are faster than others; some modes provide
settling for signals that may contain transients. The best triggering mode to use
will depend upon the measurement requirements (repeatability, accuracy and
speed).
Trigger Event
The Test Set starts a measurement cycle when a valid Trigger Event is received. A
Trigger Event is analogous to telling the Test Set to “start the measurement now.”
There are three commands that can be used to issue a Trigger Event to the Test Set
through HP-IB:
•
A Group Execute Trigger Command (GET) as defined by IEEE 488.1-1987
•
A Trigger Common Command (*TRG) as defined by IEEE 488.2-1987
•
A :TRIGger:IMMediate Test Set command.
All three commands are equivalent and have the same effect when received by the
Test Set. The Test Set responds to the three commands by triggering all currently
active measurements. A measurement is defined as active if
1. it is on the currently displayed screen
2. it is in the ON state
From a programming perspective this means that the screen which contains the
measurement of interest must be made available using the DISPlay command and
that the measurement STATe must be ON.
Содержание 8921A
Страница 14: ...14 ...
Страница 17: ...Contents 17 3 HP IB Command Guidelines Sequential and Overlapped Commands 72 Guidelines for Operation 73 ...
Страница 24: ...Contents 24 9 Error Messages ...
Страница 25: ...Contents 25 Index 593 ...
Страница 26: ...Contents 26 ...
Страница 27: ...27 1 Using HP IB ...
Страница 58: ...58 Chapter 1 Using HP IB Remote Local Modes ...
Страница 59: ...59 2 Methods For Reading Measurement Results ...
Страница 71: ...71 3 HP IB Command Guidelines ...
Страница 94: ...94 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 95: ...95 4 HP IB Commands ...
Страница 100: ...100 Adjacent Channel Power ACP ...
Страница 150: ...150 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 173: ...173 System System SYSTem ERRor Returns integer value quoted string ...
Страница 236: ...236 Triggering Measurements ...
Страница 237: ...237 5 Advanced Operations ...
Страница 350: ...350 Chapter 6 Memory Cards Mass Storage Using Memory Cards Figure 20 Inserting a Memory Card ...
Страница 360: ...360 Chapter 6 Memory Cards Mass Storage Using External Disk Drives ...
Страница 361: ...361 7 IBASIC Controller ...
Страница 442: ...442 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 559: ...559 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 562: ...562 Chapter 8 Programming The Call Processing Subsystem Example Programs ...
Страница 563: ...563 9 Error Messages ...
Страница 592: ...592 ...