418
Chapter 7, IBASIC Controller
PROGram Subsystem
:NUMber <varname>{,<nvalues>}
The :NUMBer command is used to set, from
an external controller, the value of numeric variables or arrays in an IBASIC
program in the Test Set. <varname> is the name of an existing numeric variable or
array, and can be sent as either character data (<varname> not enclosed in quotes)
or string data (<varname> enclosed in quotes). <nvalues> is a list of comma-
separated <numeric_values> which are used to set the value of <varname>.
NOTE:
If the variable name <var_name> is longer than 12 characters it must be sent as string data
(<var_name> enclosed in quotes). For example, OUTPUT 714;"PROG:NUMB
’Var_name’,10".
Attempting to send a <var_name> longer than 12 characters as character data
(<var_name> not enclosed in quotes) will generate the following error:
HP-IB Error: -112 Program mnemonic too long
.
If an attempt is made to set the value of a numeric variable or array and no
IBASIC program is in the Test Set an
IBASIC Error: -282 Illegal
program name
is generated. If an attempt is made to set the value of a numeric
variable or array and the numeric variable specified in <varname> does not exist
in the program an
IBASIC Error: -283 Illegal variable name
is
generated. If the specified numeric variable cannot hold all of the specified
<numeric_values> an
IBASIC Error: -108 Parameter not allowed
is
generated.
Syntax
PROGram[:SELected]:NUMBer <varname>{,<nvalues>}
Example setting the value of a simple variable
OUTPUT 714;"PROGram:SELected:NUMBer Variable,15"
or
OUTPUT 714;"PROG:NUMB Variable,15"
Example setting the value of a one dimensional array [Array(5)] with 6 elements
OUTPUT 714;"PROGram:SELected:NUMBer Array,0,1,2,3,4,5"
or
OUTPUT 714;"PROG:NUMB Array,0,1,2,3,4,5"
NOTE:
Individual array elements cannot be set with the :NUMBer command.
Содержание 8921A
Страница 14: ...14 ...
Страница 17: ...Contents 17 3 HP IB Command Guidelines Sequential and Overlapped Commands 72 Guidelines for Operation 73 ...
Страница 24: ...Contents 24 9 Error Messages ...
Страница 25: ...Contents 25 Index 593 ...
Страница 26: ...Contents 26 ...
Страница 27: ...27 1 Using HP IB ...
Страница 58: ...58 Chapter 1 Using HP IB Remote Local Modes ...
Страница 59: ...59 2 Methods For Reading Measurement Results ...
Страница 71: ...71 3 HP IB Command Guidelines ...
Страница 94: ...94 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 95: ...95 4 HP IB Commands ...
Страница 100: ...100 Adjacent Channel Power ACP ...
Страница 150: ...150 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 173: ...173 System System SYSTem ERRor Returns integer value quoted string ...
Страница 236: ...236 Triggering Measurements ...
Страница 237: ...237 5 Advanced Operations ...
Страница 350: ...350 Chapter 6 Memory Cards Mass Storage Using Memory Cards Figure 20 Inserting a Memory Card ...
Страница 360: ...360 Chapter 6 Memory Cards Mass Storage Using External Disk Drives ...
Страница 361: ...361 7 IBASIC Controller ...
Страница 442: ...442 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 559: ...559 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 562: ...562 Chapter 8 Programming The Call Processing Subsystem Example Programs ...
Страница 563: ...563 9 Error Messages ...
Страница 592: ...592 ...