Error
−
1311
RF power loss indicates loss of Voice Channel.
When any Call Processing Subsystem screen is displayed (except the
ANALOG MEAS
screen) and the Call Processing Subsystem is in the
Connect
state, the host firmware
constantly monitors the mobile station’s transmitted carrier power. If the power falls
below 0.0005 Watts the simulated Base Station will terminate the call and return to the
Active
state. This error message is displayed if the host firmware has detected that the
mobile station’s carrier power has fallen below the 0.0005 Watts threshold. The call is
dropped and the Call Processing Subsystem returns to the
Active
state.
NOTE:
In order to ensure that the host firmware makes the correct decisions regarding the presence
of the mobile stations’s RF carrier, the Test Set’s RF power meter should be zeroed when
first entering the Call Processing Subsystem (that is - the first time the Call Processing
Subsystem is selected during a measurement session). Failure to zero the power meter can
result in erroneous RF power measurements. See “Conditioning The Test Set For Call
Processing” in the HP 8920 User’s Guide for information on zeroing the RF Power meter
manually or <blue>"Conditioning the Test Set for Call Processing" on page 452 of this
manual for information on zeroing the RF Power meter programmatically.
Error
−
1312
Data from RVC contains invalid bits in word [1,2,3].
Indicates that the decoded data received on the reverse voice channel contains invalid bits
in word 1 and/or word 2 and/or word 3. The raw decoded data is displayed in hexadecimal
format in the top right-hand portion of the
CALL CONTROL
screen. Raw decoded data is
only displayed when the
CALL CONTROL
screen
Display
field is set to
Data.
Error
−
1313
Timeout occurred while in Maintenance state.
Indicates that the simulated Base Station’s internal timer expired before the mobile station
was taken out of the maintenance state. The internal timer is set to 20 seconds when the
maintenance order is sent to the mobile station.
Error
−
1314
Alert attempted while not in Maintenance or Connect state.
Indicates that an attempt was made to send an Alert order to the mobile station when the
Call Processing Subsystem was not in the Maintenance state or Active state.
Содержание 8921A
Страница 14: ...14 ...
Страница 17: ...Contents 17 3 HP IB Command Guidelines Sequential and Overlapped Commands 72 Guidelines for Operation 73 ...
Страница 24: ...Contents 24 9 Error Messages ...
Страница 25: ...Contents 25 Index 593 ...
Страница 26: ...Contents 26 ...
Страница 27: ...27 1 Using HP IB ...
Страница 58: ...58 Chapter 1 Using HP IB Remote Local Modes ...
Страница 59: ...59 2 Methods For Reading Measurement Results ...
Страница 71: ...71 3 HP IB Command Guidelines ...
Страница 94: ...94 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 95: ...95 4 HP IB Commands ...
Страница 100: ...100 Adjacent Channel Power ACP ...
Страница 150: ...150 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 173: ...173 System System SYSTem ERRor Returns integer value quoted string ...
Страница 236: ...236 Triggering Measurements ...
Страница 237: ...237 5 Advanced Operations ...
Страница 350: ...350 Chapter 6 Memory Cards Mass Storage Using Memory Cards Figure 20 Inserting a Memory Card ...
Страница 360: ...360 Chapter 6 Memory Cards Mass Storage Using External Disk Drives ...
Страница 361: ...361 7 IBASIC Controller ...
Страница 442: ...442 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 559: ...559 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 562: ...562 Chapter 8 Programming The Call Processing Subsystem Example Programs ...
Страница 563: ...563 9 Error Messages ...
Страница 592: ...592 ...