77
Chapter 3, HP-IB Command Guidelines
Guidelines for Operation
Specifying Units-of-Measure for Settings and Measurement Results
Numeric settings and measurement results in the Test Set can be displayed using
one or more units-of-measure (V, mV, mV, Hz, kHz, MHz…). When operating
the Test Set manually, the units-of-measure can be easily changed to display
measurement results and field settings in the most convenient format. HP-IB
operation is similar to manual operation in that the units-of-measure used to
display numeric data can be programmatically changed to the most convenient
form.
NOTE:
When querying measurements or settings through HP-IB, the Test Set always returns numeric
values in HP-IB Units or Attribute Units, regardless of the current Display Units setting. Refer
to
"HP-IB Units (UNITs)" on page 80
and
"Attribute Units (AUNits)" on page 83
for
further information.
There are three sets of units-of-measure used in the Test Set: Display Units,
HP-IB Units, and Attribute Units. Writing correct HP-IB programs requires an
understanding of how the Test Set deals with these different sets of units-of-
measure.
Display Units (DUNits)
Display Units are the units-of-measure used by the Test Set to display numeric
data (field settings and measurement results) on the front-panel CRT display. For
example, the RF Generator’s frequency can be displayed in Hz, kHz, MHz and
GHz. Similarly, the measured TX Frequency can be displayed in Hz, kHz, MHz
and GHz.
When evaluating an entered value for a numeric field, the Test Set interprets the
data it receives in terms of the Display Units currently set. For example, if the
Display Units for the
RF Gen Freq
field are set to GHz and the operator tries to
enter 500 into the field, an
Input value out of range
error is generated
since the Test Set interpreted the value as 500 GHz which is outside the valid
frequency range of the Test Set.
Содержание 8921A
Страница 14: ...14 ...
Страница 17: ...Contents 17 3 HP IB Command Guidelines Sequential and Overlapped Commands 72 Guidelines for Operation 73 ...
Страница 24: ...Contents 24 9 Error Messages ...
Страница 25: ...Contents 25 Index 593 ...
Страница 26: ...Contents 26 ...
Страница 27: ...27 1 Using HP IB ...
Страница 58: ...58 Chapter 1 Using HP IB Remote Local Modes ...
Страница 59: ...59 2 Methods For Reading Measurement Results ...
Страница 71: ...71 3 HP IB Command Guidelines ...
Страница 94: ...94 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 95: ...95 4 HP IB Commands ...
Страница 100: ...100 Adjacent Channel Power ACP ...
Страница 150: ...150 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 173: ...173 System System SYSTem ERRor Returns integer value quoted string ...
Страница 236: ...236 Triggering Measurements ...
Страница 237: ...237 5 Advanced Operations ...
Страница 350: ...350 Chapter 6 Memory Cards Mass Storage Using Memory Cards Figure 20 Inserting a Memory Card ...
Страница 360: ...360 Chapter 6 Memory Cards Mass Storage Using External Disk Drives ...
Страница 361: ...361 7 IBASIC Controller ...
Страница 442: ...442 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 559: ...559 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 562: ...562 Chapter 8 Programming The Call Processing Subsystem Example Programs ...
Страница 563: ...563 9 Error Messages ...
Страница 592: ...592 ...