576
Error
−
100
Command error
This code indicates only that a Command Error as defined in IEEE 488.2, 11.5.1.1.4 has
occurred.
Error
−
101
Invalid character
A syntactic element contains a character which is invalid for that type.
Error
−
102
Syntax error
An unrecognized command or data type was encountered; for example, a string value was
received when the device does not accept strings.
Error
−
103
Invalid separator
The parser was expecting a separator and encountered an illegal character. For example,
the colon used to separate the FREQ and AMPL commands should be omitted in the
following command:
RFG:FREQ 850 MHZ:;AMPL
−
35
Error
−
104
Data type error
The parser recognized a data element different than one allowed. For example, numeric or
string data was expected but block data was encountered.
Error
−
105
GET not allowed
A Group Execute Trigger was received within a program message (see IEEE 488.2, 7.7).
Error
−
108
Parameter not allowed
More parameters were received than expected for the header. For example, the *ESE
common command only accepts one parameter; receiving *ESE 36,1 is not allowed.
Error
−
109
Missing parameter
Fewer parameters were received than required for the header. For example, the *ESE
common command requires one parameter; receiving *ESE is not allowed.
Содержание 8921A
Страница 14: ...14 ...
Страница 17: ...Contents 17 3 HP IB Command Guidelines Sequential and Overlapped Commands 72 Guidelines for Operation 73 ...
Страница 24: ...Contents 24 9 Error Messages ...
Страница 25: ...Contents 25 Index 593 ...
Страница 26: ...Contents 26 ...
Страница 27: ...27 1 Using HP IB ...
Страница 58: ...58 Chapter 1 Using HP IB Remote Local Modes ...
Страница 59: ...59 2 Methods For Reading Measurement Results ...
Страница 71: ...71 3 HP IB Command Guidelines ...
Страница 94: ...94 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 95: ...95 4 HP IB Commands ...
Страница 100: ...100 Adjacent Channel Power ACP ...
Страница 150: ...150 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 173: ...173 System System SYSTem ERRor Returns integer value quoted string ...
Страница 236: ...236 Triggering Measurements ...
Страница 237: ...237 5 Advanced Operations ...
Страница 350: ...350 Chapter 6 Memory Cards Mass Storage Using Memory Cards Figure 20 Inserting a Memory Card ...
Страница 360: ...360 Chapter 6 Memory Cards Mass Storage Using External Disk Drives ...
Страница 361: ...361 7 IBASIC Controller ...
Страница 442: ...442 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 559: ...559 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 562: ...562 Chapter 8 Programming The Call Processing Subsystem Example Programs ...
Страница 563: ...563 9 Error Messages ...
Страница 592: ...592 ...