399
Chapter 7, IBASIC Controller
Method #3. Developing Programs Using Word Processor on a PC (Least Preferred)
Transferring Programs from the Word Processor to the Test Set
For short (less than 100 lines) programs, use an ASCII file transfer utility on the
PC to send the program, one line at a time, down to the Test Set over RS-232
directly into the IBASIC Command Line field. The Test Set must be configured to
receive serial ASCII characters by positioning the Test Set cursor at the IBASIC
Command Line field as explained under
"Method #2. Developing Programs on the
Test Set Using the IBASIC EDIT Mode" on page 392
. With this setup, when ASCII
characters are received they are sent to the IBASIC Command Line field. When a
carriage return / line feed is received, the Test Set will parse the line into the
IBASIC program memory. Each line takes about two seconds to scroll in and be
parsed. This becomes very time consuming for long programs. An alternative for
longer programs is discussed later in this section.
To start the transfer process make sure there is no program in the Test Set’s
IBASIC RAM memory by executing a SCRATCH command from the IBASIC
Command Line.
The following example shows how to transfer a short program (<100 lines) using
Microsoft Windows Terminal.
1. Make sure the Test Set cursor is in the upper left of the IBASIC Command Line field.
2. Select the
Terminal application
in the Accessories Group. Set it up as de-
scribed in earlier in this chapter.
3. Select the following:
Settings
Text Transfers
Flow Control: Line at a Time
Delay Between Lines: 25/10 Sec
Word Wrap
Outgoing Text at Column: Off.
4. Select the following:
Transfers
Send Text File
Following CR:
Strip LF selected
Append LF not selected.
5. Select the text file to be transferred and begin the transfer by selecting (OK).
Содержание 8921A
Страница 14: ...14 ...
Страница 17: ...Contents 17 3 HP IB Command Guidelines Sequential and Overlapped Commands 72 Guidelines for Operation 73 ...
Страница 24: ...Contents 24 9 Error Messages ...
Страница 25: ...Contents 25 Index 593 ...
Страница 26: ...Contents 26 ...
Страница 27: ...27 1 Using HP IB ...
Страница 58: ...58 Chapter 1 Using HP IB Remote Local Modes ...
Страница 59: ...59 2 Methods For Reading Measurement Results ...
Страница 71: ...71 3 HP IB Command Guidelines ...
Страница 94: ...94 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 95: ...95 4 HP IB Commands ...
Страница 100: ...100 Adjacent Channel Power ACP ...
Страница 150: ...150 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 173: ...173 System System SYSTem ERRor Returns integer value quoted string ...
Страница 236: ...236 Triggering Measurements ...
Страница 237: ...237 5 Advanced Operations ...
Страница 350: ...350 Chapter 6 Memory Cards Mass Storage Using Memory Cards Figure 20 Inserting a Memory Card ...
Страница 360: ...360 Chapter 6 Memory Cards Mass Storage Using External Disk Drives ...
Страница 361: ...361 7 IBASIC Controller ...
Страница 442: ...442 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 559: ...559 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 562: ...562 Chapter 8 Programming The Call Processing Subsystem Example Programs ...
Страница 563: ...563 9 Error Messages ...
Страница 592: ...592 ...