231
Triggering Measurements
Default Trigger Mode
The Trigger mode is set to FULL SETTling and REPetitive RETRiggering
whenever
•
the Test Set is powered on
•
the PRESET key is selected
•
the Test Set is put into LOCAL mode
•
the Test Set is reset using the *RST command
•
the Test Set is put in remote mode and no other trigger mode is set
Local/Remote Triggering Changes
Local To Remote Transitions
The Test Set switches from Local to Remote mode upon receipt of the Remote
message (REN bus line true and Test Set is addressed to listen). No instrument
settings are changed by the transition from Local to Remote mode, but triggering
is set to the state it was last set to in Remote mode (if no previous setting, the
default is FULL SETTling and REPetitive RETRiggering).
When the Test Set makes a transition from local to remote mode, all currently
active measurements are flagged as invalid causing any currently available
measurement results to become unavailable. If the HP-IB trigger mode is
:RETR REP then a new measurement cycle is started and measurement results
will be available for all active measurements when valid results have been
obtained. If the HP-IB trigger mode is :RETR SING then a measurement cycle
must be started by issuing a trigger event. Refer to
for more information.
Remote To Local Transitions
The Test Set switches from Remote to Local mode upon receipt of the Local
message (Go To Local bus message is sent and Test Set is addressed to listen) or
receipt of the Clear Lockout/Set Local message (REN bus line false). No
instrument settings are changed by the transition from Remote to Local mode, but
triggering is reset to FULL SETTling and REPetitive RETRiggering.
Содержание 8921A
Страница 14: ...14 ...
Страница 17: ...Contents 17 3 HP IB Command Guidelines Sequential and Overlapped Commands 72 Guidelines for Operation 73 ...
Страница 24: ...Contents 24 9 Error Messages ...
Страница 25: ...Contents 25 Index 593 ...
Страница 26: ...Contents 26 ...
Страница 27: ...27 1 Using HP IB ...
Страница 58: ...58 Chapter 1 Using HP IB Remote Local Modes ...
Страница 59: ...59 2 Methods For Reading Measurement Results ...
Страница 71: ...71 3 HP IB Command Guidelines ...
Страница 94: ...94 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 95: ...95 4 HP IB Commands ...
Страница 100: ...100 Adjacent Channel Power ACP ...
Страница 150: ...150 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 173: ...173 System System SYSTem ERRor Returns integer value quoted string ...
Страница 236: ...236 Triggering Measurements ...
Страница 237: ...237 5 Advanced Operations ...
Страница 350: ...350 Chapter 6 Memory Cards Mass Storage Using Memory Cards Figure 20 Inserting a Memory Card ...
Страница 360: ...360 Chapter 6 Memory Cards Mass Storage Using External Disk Drives ...
Страница 361: ...361 7 IBASIC Controller ...
Страница 442: ...442 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 559: ...559 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 562: ...562 Chapter 8 Programming The Call Processing Subsystem Example Programs ...
Страница 563: ...563 9 Error Messages ...
Страница 592: ...592 ...