Chapter 7, IBASIC Controller
Interfacing to the IBASIC Controller using Serial Ports
Serial Port 9 Configuration
and the following paragraphs describe how to configure
Serial Port 9 for communications with an external PC or terminal. Implications of
the various choices are discussed.
1. Under the
To Screen
menu, select
More
, then select IO CONFIG.
2. The I/O CONFIGURE screen will be displayed.
3. Set the
Serial Baud Rate
,
Parity
,
Data Length
,
Stop Length
,
Rcv
Pace
and
Xmt Pace
fields to match your PC or terminal settings. The recommended
settings are shown in
. These settings will be retained by the Test
Set. They will not change if the PRESET key is pressed, if the Test Set receives a *RST
Common Command, or the power is turned on and off.
4. Set the
Serial In
field to
Inst
. This routes Serial Port 9 to the
IBASIC Command
Line
field. Characters typed on the external PC or terminal will now appear in the
IBASIC Command Line
.
5. Set the
IBASIC Echo
field to
ON
. This will cause IBASIC character output from
commands (such as LIST, PRINT or DISPLAY) or error messages to echo characters
to Serial Port 9 (the characters will in turn show up on the external PC or terminal
screen). This will allow program listings and syntax error messages to be seen on the
external PC or terminal.
6. Another method which can be used to output characters to the external PC or terminal
is to execute the IBASIC command, PRINTER IS 9. This causes IBASIC to direct all
print output to Select Code 9. Select Code 9 is the Test Set’s Serial Port 9. Select Code
1 is the Test Set’s CRT. Select Code 1 is also the default address for the PRINTER IS
command, so all program printer output defaults to the Test Set’s CRT (unless changed
with the PRINTER IS command).
7. Set the
Inst Echo
field to
ON
. This will cause characters to be echoed back to the
external PC or terminal as they are received at Serial Port 9. If the echo feature of the
external PC or terminal is also enabled all the characters sent to the Test Set will be
displayed twice on the external PC or terminal. Enable echo on only one device, either
the Test Set or the external PC or terminal.
Содержание 8921A
Страница 14: ...14 ...
Страница 17: ...Contents 17 3 HP IB Command Guidelines Sequential and Overlapped Commands 72 Guidelines for Operation 73 ...
Страница 24: ...Contents 24 9 Error Messages ...
Страница 25: ...Contents 25 Index 593 ...
Страница 26: ...Contents 26 ...
Страница 27: ...27 1 Using HP IB ...
Страница 58: ...58 Chapter 1 Using HP IB Remote Local Modes ...
Страница 59: ...59 2 Methods For Reading Measurement Results ...
Страница 71: ...71 3 HP IB Command Guidelines ...
Страница 94: ...94 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 95: ...95 4 HP IB Commands ...
Страница 100: ...100 Adjacent Channel Power ACP ...
Страница 150: ...150 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 173: ...173 System System SYSTem ERRor Returns integer value quoted string ...
Страница 236: ...236 Triggering Measurements ...
Страница 237: ...237 5 Advanced Operations ...
Страница 350: ...350 Chapter 6 Memory Cards Mass Storage Using Memory Cards Figure 20 Inserting a Memory Card ...
Страница 360: ...360 Chapter 6 Memory Cards Mass Storage Using External Disk Drives ...
Страница 361: ...361 7 IBASIC Controller ...
Страница 442: ...442 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 559: ...559 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 562: ...562 Chapter 8 Programming The Call Processing Subsystem Example Programs ...
Страница 563: ...563 9 Error Messages ...
Страница 592: ...592 ...