205
Equivalent Front-Panel Key Commands
LOCAL
The LOCAL key returns the Test Set to local, front-panel control. The Test Set
returns to Local operation (full front-panel control) when either the Go To Local
(GTL) bus command is received, the front-panel LOCAL key is pressed or the
REN line goes false. When the Test Set returns to local mode the output signals
and internal settings remain unchanged, except that triggering is reset to
TRIG:MODE:SETT FULL;RETR REP. The LOCAL key will not function if the
Test Set is in the local lockout mode.
MEAS RESET
The MEAS RESET key clears the measurement history for all of the Test Set’s
measurement algorithms: Averaging (AVG key, Spectrum Analyzer trace
averaging), Measurement limit checking (HI LIMIT and LO LIMIT keys), Peak
Hold (AF Analyzer peak hold detectors, Spectrum Analyzer trace peak hold),
autotuning and autoranging, and re-starts all active measurements. The HP-IB
commands :MEASure:RESet are used to select this function programmatically.
Syntax
:MEASure:RESet
Example
OUTPUT 714;":MEAS:RES"
This resets all of the active measurements in the Test Set.
PRESET
The PRESET key resets the Test Set to its power-up state. The IEEE 488.2
Common Command *RST is used to select this function programmatically.
Syntax
*RST
Example
OUTPUT 714;"*RST"
This resets the Test Set to its power-up state.
Содержание 8921A
Страница 14: ...14 ...
Страница 17: ...Contents 17 3 HP IB Command Guidelines Sequential and Overlapped Commands 72 Guidelines for Operation 73 ...
Страница 24: ...Contents 24 9 Error Messages ...
Страница 25: ...Contents 25 Index 593 ...
Страница 26: ...Contents 26 ...
Страница 27: ...27 1 Using HP IB ...
Страница 58: ...58 Chapter 1 Using HP IB Remote Local Modes ...
Страница 59: ...59 2 Methods For Reading Measurement Results ...
Страница 71: ...71 3 HP IB Command Guidelines ...
Страница 94: ...94 Chapter 3 HP IB Command Guidelines Guidelines for Operation ...
Страница 95: ...95 4 HP IB Commands ...
Страница 100: ...100 Adjacent Channel Power ACP ...
Страница 150: ...150 Display DISPlay space TDMA test TESTs TFReq THLP TIBasic TMAKe TPARm TPRint TSEQn TSPec Returns current screen TX ...
Страница 173: ...173 System System SYSTem ERRor Returns integer value quoted string ...
Страница 236: ...236 Triggering Measurements ...
Страница 237: ...237 5 Advanced Operations ...
Страница 350: ...350 Chapter 6 Memory Cards Mass Storage Using Memory Cards Figure 20 Inserting a Memory Card ...
Страница 360: ...360 Chapter 6 Memory Cards Mass Storage Using External Disk Drives ...
Страница 361: ...361 7 IBASIC Controller ...
Страница 442: ...442 Chapter 7 IBASIC Controller The TESTS Subsystem ...
Страница 559: ...559 Chapter 8 Programming The Call Processing Subsystem Example Programs 11280 SUBEND 11290 ...
Страница 562: ...562 Chapter 8 Programming The Call Processing Subsystem Example Programs ...
Страница 563: ...563 9 Error Messages ...
Страница 592: ...592 ...