3-11
Normalization
Adjusting the Power Monitor
5
Enter 0.0 in the
Power constant
field. Click
OK
.
6
Press the
Local
key on the TLS (tuneable laser source) to put it in local
operation mode.
7
On the TLS, set the wavelength to 1558.5 nm and the power to 0 dBm.
8
Connect a short fiber-optic cable from the
SOURCE OUTPUT
connector to an
optical power meter. Press the
Active
key on the tuneable laser source.
9
Measure the optical power at 1558.5 nm on the power meter and record the
reading P
1
.
P
1
: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ____________ dBm
10
Disconnect the fiber-optic cable from the power meter and connect it to the
RECEIVER INPUT
port on the HP 83427A test set.
11
Click the
Power Monitor
button on the main window and record the power
reading P
2
.
P
2
: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ____________ dBm
12
Calculate the difference using the following equation:
∆
P = P
1
– P
2
P
∆
: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ____________ dBm
For example:
∆
P = (–12 dBm) – (–8 dBm)
∆
P = –4 dBm
13
Close the
Power Monitor
window.
14
Choose
Power Monitor Settings
from the
Service
menu.
15
In the
Power Monitor Setup
dialog box, enter the following values:
Power constant:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ∆
P (as calculated above)
Under range limit: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –43 dBm
Insertion loss reference: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . –15 dBm
TLS power:. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 dBm
TLS wavelength: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1558.5 nm
16
Click
OK
.
17
Click the
Power Monitor
button on the main window. Verify that the displayed
power is within ±0.5 dB of value P
1
.
18
Click the
Set zero
button from within the
Power Monitor
window.
19
Close the
Power Monitor
window. The TLS automatically reverts to system
Содержание 86037B
Страница 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Страница 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Страница 72: ...3 16 Normalization Setting Normalization Preferences ...
Страница 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Страница 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Страница 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Страница 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Страница 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Страница 120: ...4 48 Remote Control Output Parameters ...
Страница 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Страница 165: ...6 7 Installation Installation 2 3 4 5 1 8 7 Fan Pack 6 6 Power Distribution Assembly 1 7 Power Distribution Assembly 2 ...
Страница 170: ...6 12 Installation Installation ...
Страница 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Страница 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Страница 190: ...7 20 Reference System Options ...
Страница 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Страница 198: ...8 8 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 204: ...Index 6 Index ...