5-6
Maintenance
Troubleshooting Common Problems
1
Inquire about recent repairs or changes to the system.
Often, this will help pinpoint current problems even if the system was operat-
ing correctly after the repair or change to the system.
2
Setup or reinstall the HP 86037B software to determine whether the problem
is a hardware or a software problem.
Refer to “Reinstalling the System
.
for the correct addresses for the computer and
the HP 86037B Chromatic Dispersion Test System. If reinstalling the software
does not solve the problem, the problem may not be a software error.
3
Check that the system components are wired correctly.
4
Remove, clean, and reinstall all optical connections.
5
Ensure that all cables are secure.
6
Check for a 50
Ω
load at the back of each TLS.
To check the TLS
W A R N I N G
Prevent any accidental exposure to the laser beam. Prior to
disconnecting any optical fibers on the TLS, ensure that the TLS is
turned off.
1
Turn off the power on the TLS. Failure to comply can cause personal injury.
2
Remove the cable connecting the TLS to the test set.
3
Connect a short fiber-optic cable between the optical power meter and the
TLS. The distance between the optical power meter and the TLS should be
kept to a minimum.
4
Turn on the TLS.
5
If the TLS power measures low, the TLS needs repair.
6
Turn off the power on the TLS. Failure to comply can cause personal injury.
7
Remove the fiber-optic cable connecting the TLS and the optical power meter.
8
Reconnect the original fiber-optic cable that you removed in
Step 2
.
9
Connect the open end of the fiber-optic cable to the optical power meter.
10
Turn on the TLS.
11
Test the TLS power. If the power measures low, replace the fiber-optic cable.
Содержание 86037B
Страница 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Страница 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Страница 72: ...3 16 Normalization Setting Normalization Preferences ...
Страница 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Страница 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Страница 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Страница 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Страница 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Страница 120: ...4 48 Remote Control Output Parameters ...
Страница 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Страница 165: ...6 7 Installation Installation 2 3 4 5 1 8 7 Fan Pack 6 6 Power Distribution Assembly 1 7 Power Distribution Assembly 2 ...
Страница 170: ...6 12 Installation Installation ...
Страница 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Страница 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Страница 190: ...7 20 Reference System Options ...
Страница 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Страница 198: ...8 8 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 204: ...Index 6 Index ...