5-11
Maintenance
Verification Tests
3
After the
Preset
is complete, the following parameters appear on the network
analyzer’s display:
CH1, S11, log MAG, 10dB, REF 0dB, START 30kHz, STOP 3GHz
(or
6GHz for Option 006).
If the self test fails, refer to the HP 8753E Network Analyzer service manual
for instructions on testing the individual components of the network analyzer,
or call your HP service center.
To run the the network analyzer operator’s check
The operator’s check determines that the source is phase locked across the
entire frequency range and that all three samplers are functioning properly.
Figure 5-3. Operator’s Check for the HP 8753 Network Analyzer
1
Disconnect all devices, peripherals, and accessories (including adapters and
limiters) from the HP 8753 network analyzer.
2
Switch on the network analyzer.
3
Press the
Preset
button on the network analyzer.
4
Press the
System
button on the network analyzer.
5
On the network analyzer:
a
Press the softkey corresponding to the
Service
menu.
b
Press the softkey corresponding to the
Tests
menu.
c
Enter,
21
, on the network analyzer keypad.
d
Press the, x1, button on the network analyzer.
e
Press the softkey corresponding to the
External Tests
function.
6
Press the
Execute Test
key on the network analyzer.
7
At the prompt, connect the equipment as shown in
, with power to
Table 5-5. Equipment Required
Equipment
Model Number
20 dB attenuator
HP 8491A Option 020
RF cable set
HP 11851B
Two-way power schooner
HP 11667A Option 001
Содержание 86037B
Страница 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Страница 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Страница 72: ...3 16 Normalization Setting Normalization Preferences ...
Страница 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Страница 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Страница 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Страница 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Страница 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Страница 120: ...4 48 Remote Control Output Parameters ...
Страница 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Страница 165: ...6 7 Installation Installation 2 3 4 5 1 8 7 Fan Pack 6 6 Power Distribution Assembly 1 7 Power Distribution Assembly 2 ...
Страница 170: ...6 12 Installation Installation ...
Страница 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Страница 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Страница 190: ...7 20 Reference System Options ...
Страница 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Страница 198: ...8 8 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 204: ...Index 6 Index ...