8-4
Specifications and Regulatory Information
Specifications
Optical power range for RECEIVER INPUT
–3 to –43 dBm (optical)
Modulation frequency range
100 MHz to 2.5 GHz
Dispersion measurement range
0.1 ps/nm to Dmax
c
Group delay uncertainty
d
0.5 ps
(0.05 ps characteristic)
Zero dispersion wavelength repeatability
e
0.05 nm
(0.005 nm characteristic)
Repeatability of dispersion slope at zero dispersion
wavelength
e
0.002 ps/(nm2 × km) (characteristic)
Zero dispersion wavelength accuracy
f
with wavelength meter (With HP 8168F)
without wavelength meter (With HP 8168F)
without wavelength meter (With HP 8168D)
Relative amplitude accuracy
±0.05 nm
±0.15 nm
±0.25 nm
±0.25 dB (characteristic)
Measurement time (characteristic)
g
with wavelength meter
without wavelength meter
3 seconds per point
2 seconds per point
a. Refer to power specification for tunable laser source for other wavelength ranges.
b. System dynamic range is calculated by subtracting the test set loss and receiver sensitivity from the tunable laser source’s output power.
c. Maximum dispersion, Dmax (in ns/nm), is given by 25/Dl, where Dl is the wavelength step size in nm.
d. Based on noise floor measurements, full output power, reference wavelength on, no averaging, and high substitute
e. One standard deviation for 20 measurements on the same spool of 10 km dispersion-shifted fiber for the wavelength range from 1515 to
1580 nm with power into RECEIVER INPUT at -43 dBm (optical). Parameters refer to curve fit traces using the five-term Sellmeier fit.
f. These values are based on measurements of NIST Standard Reference Material 2524, Serial No. 0001, using a quadratic fit to 65 point
spaced at 1 nm.
g. This calculation was done for 65 points on a standard spool of 10 km fiber using normal sensitivity. Any change of sensitivity will result in
a corresponding change in time. As an example, 10 runs of standard fiber with 65 points per run takes about one hour.
Table 8-1. Chromatic Dispersion Measurement Specifications (Continued)
Measurement
Specification
Содержание 86037B
Страница 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Страница 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Страница 72: ...3 16 Normalization Setting Normalization Preferences ...
Страница 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Страница 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Страница 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Страница 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Страница 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Страница 120: ...4 48 Remote Control Output Parameters ...
Страница 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Страница 165: ...6 7 Installation Installation 2 3 4 5 1 8 7 Fan Pack 6 6 Power Distribution Assembly 1 7 Power Distribution Assembly 2 ...
Страница 170: ...6 12 Installation Installation ...
Страница 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Страница 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Страница 190: ...7 20 Reference System Options ...
Страница 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Страница 198: ...8 8 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 204: ...Index 6 Index ...