1-12
Quick Start
Measuring Grating Devices
Without the benefit of curve fitting and since dispersion is the derivative of the
group delay, the effect of any noise is exaggerated. For example, large
amounts of noise will be measured at wavelengths in a filter’s cutoff region.
5
Use a 2 GHz RF modulation frequency with a small wavelength step size. Step
sizes down to 1 pm are possible. To set the step size,
.
Setting the RF modulation frequency
When selecting the RF modulation frequency, you’ll want to select the highest frequency
possible without introducing 360
°
phase shifts. This requires that you know the approxi-
mate group delay variation (between measurement wavelength steps) expected on the
device that you are measuring.
Set the RF modulation frequency to a value which has a period that is at least twice the
anticipated group delay between wavelength steps.
For example, when measuring a grating device that has a variation in group delay of as
much as 500 ps between wavelength steps, set the RF modulation frequency to 1 GHz or
less. If you specify a wavelength step size of 2 pm to 100 pm, it is unlikely that any but the
fastest gratings (such as tuning filters or dispersion compensation gratings) would require
less than the default 2 GHz frequency.
RF modulation frequency
1
2 group delay between wavelength steps
(
)
-----------------------------------------------------------------------------------------------------------------
≤
Содержание 86037B
Страница 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Страница 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Страница 72: ...3 16 Normalization Setting Normalization Preferences ...
Страница 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Страница 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Страница 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Страница 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Страница 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Страница 120: ...4 48 Remote Control Output Parameters ...
Страница 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Страница 165: ...6 7 Installation Installation 2 3 4 5 1 8 7 Fan Pack 6 6 Power Distribution Assembly 1 7 Power Distribution Assembly 2 ...
Страница 170: ...6 12 Installation Installation ...
Страница 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Страница 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Страница 190: ...7 20 Reference System Options ...
Страница 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Страница 198: ...8 8 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 204: ...Index 6 Index ...