5-5
Maintenance
Troubleshooting Common Problems
data. Because the 5-term Sellmeier curve fit has more freedom to follow the
shape of the data, it may display multiple zero dispersion wavelengths under
some conditions.
The three types of curve fit are appropriate in different situations, as shown in
the following table.
For most optical fibers, only the 5-term Sellmeier curve can provide a good fit
over the combined (1300 and 1500 nm) wavelength ranges. A fitted 3-term
Sellmeier or Quadratic curve follow the data quite closely in one wavelength
range but diverge from the data in the other wavelength range. This is to be
expected from the capabilities of the curve fit “model equations”.
If fitted curve does not appear in relative group delay graph
❒
Refer to “If fitted curve does not follow the actual data” on page 5-4
.
Depending upon the device under test and the setup conditions, the fitted
curve may closely follow the data in one of the wavelength ranges and diverge
from the data in the other wavelength range. This is more likely to happen
when the 3-term Sellmeier or Quadratic curve fit is applied to data collected in
the combined wavelength ranges.
To perform a quick check
Common troubleshooting skills can be used to locate many problems. The fol-
lowing steps are suggested for quickly narrowing down the problem.
Table 5-1. Types of Fitted Curves
Wavelength Range
of Measured Data
Type of Fiber
Suggested Curve Fit
1300 nm
non-dispersion shifted (SMF)
3-term Sellmeier
dispersion shifted (SMF/DS)
Quadratic or 3-term
1500 nm
non-dispersion shifted (SMF)
Quadratic or 3-term
dispersion shifted (SMF/DS)
Quadratic
1300 and 1500 nm
non-dispersion shifted (SMF)
5-term Sellmeier
dispersion shifted (SMF/DS)
5-term Sellmeier
Содержание 86037B
Страница 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Страница 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Страница 72: ...3 16 Normalization Setting Normalization Preferences ...
Страница 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Страница 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Страница 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Страница 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Страница 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Страница 120: ...4 48 Remote Control Output Parameters ...
Страница 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Страница 165: ...6 7 Installation Installation 2 3 4 5 1 8 7 Fan Pack 6 6 Power Distribution Assembly 1 7 Power Distribution Assembly 2 ...
Страница 170: ...6 12 Installation Installation ...
Страница 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Страница 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Страница 190: ...7 20 Reference System Options ...
Страница 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Страница 198: ...8 8 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 204: ...Index 6 Index ...