5-26
Maintenance
Cleaning Connections for Accurate Measurements
duce a good return-loss measurement. The quality of the polish establishes
the difference between the “PC” (physical contact) and the “Super PC” con-
nectors. Most connectors today are physical contact which make glass-to-glass
connections, therefore it is critical that the area around the glass core be clean
and free of scratches. Although the major area of a connector, excluding the
glass, may show scratches and wear, if the glass has maintained its polished
smoothness, the connector can still provide a good low level return loss con-
nection.
If you test your cables and accessories for insertion loss and return loss upon
receipt, and retain the measured data for comparison, you will be able to tell in
the future if any degradation has occurred. Typical values are less than 0.5 dB
of loss, and sometimes as little as 0.1 dB of loss with high performance con-
nectors. Return loss is a measure of reflection: the less reflection the better
(the larger the return loss, the smaller the reflection). The best physically
contacting connectors have return losses better than 50 dB, although 30 to 40
dB is more common.
To Test Insertion Loss
Use an appropriate lightwave source and a compatible lightwave receiver to
test insertion loss. Examples of test equipment configurations include the fol-
lowing equipment:
• HP 71450A or HP 71451A Optical Spectrum Analyzers with Option 002 built-in
white light source.
• HP 8702 or HP 8703 Lightwave Component Analyzer system.
• HP 83420 Chromatic Dispersion Test Set with an HP 8510 Network Analyzer.
• HP 8153 Lightwave Multimeter with a source and power sensor module.
To Test Return Loss
Use an appropriate lightwave source, lightwave receiver, and lightwave cou-
pler to test return loss. Examples of test equipment configurations include the
following equipment:
• HP 8703 Lightwave Component Analyzer.
• HP 8702 Lightwave Component Analyzer with the appropriate source,
receiver, and lightwave coupler.
• HP 8504 Precision Reflectometer.
• HP 8153 Lightwave Multimeter with a source and power sensor module in
conjunction with a lightwave coupler.
• HP 81554SM Dual Source and HP 81534A Return Loss Module.
Содержание 86037B
Страница 1: ...User s Guide HP 86037B Chromatic Dispersion Test System ...
Страница 31: ...2 13 Operation The Main Window Figure 2 4 Repeat Measurement Selector ...
Страница 72: ...3 16 Normalization Setting Normalization Preferences ...
Страница 94: ...4 22 Remote Control TES Parameters Main chromatic dispersion screen ...
Страница 100: ...4 28 Remote Control TES Parameters CD and Gain Measurement Settings Dialog Box ...
Страница 102: ...4 30 Remote Control TES Parameters Length Measurement Settings Dialog Box ...
Страница 104: ...4 32 Remote Control TES Parameters Normalization Settings Dialog Box ...
Страница 106: ...4 34 Remote Control TES Parameters New Normalization Button Dialog Box ...
Страница 120: ...4 48 Remote Control Output Parameters ...
Страница 158: ...5 38 Maintenance Hewlett Packard Sales and Service Offices ...
Страница 165: ...6 7 Installation Installation 2 3 4 5 1 8 7 Fan Pack 6 6 Power Distribution Assembly 1 7 Power Distribution Assembly 2 ...
Страница 170: ...6 12 Installation Installation ...
Страница 174: ...7 4 Reference Theory of Operation Figure 7 3 HP 86037B Block Diagram Options 121 and 122 ...
Страница 175: ...7 5 Reference Theory of Operation Figure 7 4 The HP 86037B Option 114 block diagram ...
Страница 190: ...7 20 Reference System Options ...
Страница 191: ...8 Specifications 8 3 Regulatory Information 8 7 Specifications and Regulatory Information ...
Страница 198: ...8 8 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 204: ...Index 6 Index ...