5-10.
YO DELAY COMPENSATION, A54
Reference
Performance Test: Swept Frequency Accuracy
Service Section: Sweep Generator
-
YO Loop
Description
In this procedure, the programmable width of the YO kick pulse is calibrated. Then the YO delay is
adjusted so the marker position tracks from the slowest to the fastest sweep speeds.
Equipment
Spectrum Analyzer
HP 8566B
Oscilloscope
HP 1741A
Adapter, APC-3.5 (f) to Type N
(in)
HP P/N 1250-1744
(2 required for HP 8341 B, one for HP 8340B)
Cable, SMA
(in)
to SMA
(in)
HP P/N 08340-201 24
Adapter, APC 3.5 (f) to APC 3.5 (f) (required for HP 8340B)
HP P/N 5061-5311
NOTE
In the test setup, 2 APC 3.5 (f) to Type N (in) adapters are used with an
SMA semi-rigid cable. This setup is used, instead of connecting the DUT
to the spectrum analyzer with a Type N cable, to reduce power loss
through the cable.
OSCILLOSCOPE
DUT
AE4TP2
ILK ICKI
Figure 5-28.
VO Kick Pulse Test Setup
HP 8340B/41B
Adjustments
5-57