YO DELAY COMPENSATION, A54 (Cont’d)
24. Press the OUT’s [RF] key to turn RF poweron. Adjust A54R30 CGN to set the peak of the signal at
the center line of the oscilloscope screen.
25. On the OUT, press [RECALL],then enter [1]. Repeat steps 7 through 24 until no further adjust-
ment of A54R30 (CGN) and A54R32 (COFF) is necessary.
26. Disconnect all of the test equipment and reconnect the cables.
BEGINNING
OF MARKER
Figure 5-31.
Z-Axis Marker Waveform
Figure 5-32.
Amplitude Marker Waveform