5-14.
UNLEVELED RF OUTPUT ADJUSTMENTS (Cont’d)
48. Access Cal Constant 8
by
pressing
[SHIFT][GHz][8] [Hz] [SHIFT][ENTRYOFF] on the OUT.
Using the OUT’s rotary knob, adjust Cal
Constant
8
(range is
0
to
131) for optimum power across
Band 4.
49. Press [RECALL]on the OUT and iterate between registers 1
and
2 by
pressing
[1] and then [2].
Verify that the fast sweep speed
trace (register
2)
is no more
than
1 dB
below the slow sweep
speed trace (register
1)
at any point.
Also, the minimum power point on the fast sweep speed trace
should
be> 1 dB above
the maximum
leveled power specifications for the Band. If either of the
above conditions are not met, repeat steps 46 through 49.
Proceed with step 51.
50.
Perform
steps
45
and
46.
Band 2 Delay Compensation, HP 8340B and 8341B
51. Press the following key sequence on the OUT:
[CONT](If not already in continuous sweep mode)
[STARTFREQ] [6]
[.]
[9] [GHz]
[STOPFREQ] [1] [3]
[.]
[5] [GHz]
[SWEEPTIME][2] [0] [0] [msec]
(500
mS if the OUT is an HP 8341B Option 003)
[SAVE][1]
[SWEEPTIME][AUTO]
[SAVE][2]
[SINGLE]
[SAVE][3]
[RECALL][2]
Adjust the scope’s vertical position and/or sensitivity to view the
swept
display.
52. Access Cal Constant
6 by
pressing
[SHIFT][GHz][6] [Hz] [SHIFT][ENTRYOFF]
on
the
OUT.
Using the OUT’s rotary knob, adjust Cal Constant 6 (range is 0 to 131) to optimize the power
across
the entire Band.
53.
Press [RECALL] on the
OUT
and iterate between registers 1 and 2 by pressing [1] and then [2].
Verify that the fast sweep speed trace (register 2) is no more than
1 dB
below the slow sweep
speed trace (register 1) at any point. Also, the minimum power point on the fast sweep speed trace
should be
>
1 dB above the maximum leveled power specifications for Band 2. If either of the
above conditions are not met, repeat steps 52 and 54.
54. Iterate between registers 1 and 3 and verify that the single sweep trace (register 3) is no more than
1 db
below the slow sweep trace at any point. After pressing [3] on the OUT you need to press
[SINGLE]
to initiate
the single sweep. Also, the minimum power point on the single sweep trace
should be
>
1 dB above the maximum leveled power specifications for Band 2.
If
either of the
above conditions are not met, repeat steps 52 through 54.
HP 8340B/41B
Adjustments