5-14.
UNLEVELED RF OUTPUT ADJUSTMENTS (Cont’d)
3.
Press the [.~] step key to access each Cal Constant listed in the table below and preset the Cal
Constants to the value provided in the table. Be sure to terminate each entry by pressing [Hz].
Cal
Constant
Value
Cal
Constant
Value
3
100
52
0
4
100
53
0
5
100
71
1024
6
100
72
1024
7
100
73
1024
8
100
74
1024
9
1024
75
25
10
1024
76
1000
11
1024
77
—25
12
1024
78
25
50
0
80
0
51
0
4.
If
the A13 SYTM or the A24 Attenuator Driver/SRD Bias assembly has been replaced, preset the
following potentiometers (pots) to the position stated (see Figure 5-41):
Center the following:
A24R3 (X2A)
A24R4 (X2B)
Set the following fully counter-clockwise:
A24R1 (OFF A)
A24R2 (OFF B)
A24R8 (X3C)
A24R11 (X4C)
5.
If the A13 SYTM or the A28 SYTM Driver assembly has been replaced, preset the following pots to
the position stated (see Figure
5-41):
A24R6 (X3A)
A24R7 (X3B)
A24R9 X4A)
A24R10 (X4B)
A24R5 (X2C)
A24R12 (MIN)
Center:
A28R6 (OFST)
Set the following fully clockwise:
A28R13 (DRP)
A28R24 (BP2 FREQ)
Set the following fully counter-clockwise:
A28R19 (DYS)
A28R22 (DYO)
A28R28 (BP3)
A28R82 (GAIN)
A28R25 (BP3 FREQ)
A28R27 (BP2)
A28R99 (BP1)
A28R88 (KICK)
HP 8340B/41B
Adjustments
5-75