5-16.
LEVELED RF OUTPUT ADJUSTMENTS (Cont’d)
49. Use the step keys to step through Band 4. If squegging occurs, adjust A24R11 (X4C) slightly CCW
to eliminate the squegging.
NOTE
Test for squegging at power levels from maximum leveled power to —20
dBm in 5 dB increments.
50. Press the OUT’s [SHIFT] [PWRSWP] and enter the maximum ALC power level that will be a 5
dB
increment below max leveled power (i.e., 5, 0, —5). Repeat step
49 at this
power level. Enter the
next 5 dB increment and repeat step 49 until the test for squegging has been performed from
maximum leveled power to —20 dBin.
5-114
Adjustments
HP 8340B/41B