Model 8340A - Service
Incorrect Sweep
If the Marker Ramp waveform is correct, but the output of the
SWEEP GENERATOR is still incorrect, then the various gain and
attenuation stages must be checked. Set the instrument to sweep
from 2.3 GHz to 7 GHz with a sweep time of 10 msec. The voltage
at the MKR RMP test point, TP4, should be a ramp from 0 to +10
volts with a forward sweep time of 10 msec. Looking at the RST
RMP test point, TP2, should show a ramp going from O to -10 volts
in 10 msec. If this test point is stuck at ground, then check Ul3
pin 14 in the RESET AMPLIFIER 2 block. It should be a TTL HIGH
during the forward sweep of the ramp and LOW the rest of the
time.
Next look at the 20-30 SWP test point, TPB. It should be a ramp
going from O to +9.4 volts. If it is not, then go to the single
sweep mode and press the following key sequence:
[SHIFT]
[XTAL]
(leveling). Then press the
[SINGLE]
SWEEP key once. This will
cause the Marker Ramp to sweep to +10 volts and stop. Now, using
Direct I/O, write a O to channel 1 sub-channel O by pressing this
key sequence:
[SHIFT]
[GHz]
(1]
[Hz]
[SHIFT]
[MHz]
[0]
[Hz]
[SHIFT]
[KHz]
(O]
[Hz]
•
Check the digital input to Ull (pins 4-15), they should all be
LOW. TPB should be at ground.
Now write a 4095 to the same strobe using the following key
sequence:
[ 4]
(O]
[9]
[5]
[Hz]
•
This should make all the digital
inputs to Ull HIGH. The voltage at TPB should now be the opposite
of the voltage at TP2. That is, since TP2 should be at
approximately -10 volts, TPB should be at about +10 volts. By
turning the front panel RPG slowly and decreasing the number
written to this strobe from 4095 to zero, the voltage at TP8
should go smoothly from +10 volts to ground. Press
[SHIFT]
[XTAL]
to return the �nstrument to the normal operating mode.
Go to the continuous sweep mode and now check to see that the
VSWP test point, TPlO, is going from O to +9.4 volts. If not,
then check to see that the gates of Q7-Ql2 in the SWEEP WIDTH
RANGE ATTENUATOR are at the correct voltage. See the circuit
description Block
N
for the appropriate values.
Finally, if the instrument does single band sweeps correctly but
does not do multi-band ones, check U5. This can be done by going
8-375
Scans by HB9HCA and HB9FSX
Содержание 8340A
Страница 1: ...Scans by HB9HCA and HB9FSX ...
Страница 113: ...Scans by HB9HCA and HB9FSX ...
Страница 187: ...Scans by HB9HCA and HB9FSX ...
Страница 198: ...Scans by HB9HCA and HB9FSX ...
Страница 269: ...Scans by HB9HCA and HB9FSX ...
Страница 296: ...Scans by HB9HCA and HB9FSX ...
Страница 320: ...Scans by HB9HCA and HB9FSX ...
Страница 321: ...Scans by HB9HCA and HB9FSX ...
Страница 322: ...Scans by HB9HCA and HB9FSX ...
Страница 323: ...Scans by HB9HCA and HB9FSX ...
Страница 324: ...Scans by HB9HCA and HB9FSX ...
Страница 325: ...Scans by HB9HCA and HB9FSX ...
Страница 326: ...Scans by HB9HCA and HB9FSX ...