Model 8340A - Service
Band Number
YO Frequency
r
UNLEVELED
START
ow
,--
STOP
°'
OVERMOD
REMOTE
'""
FAULT
OVEN
EXT REF
UNLK
FREQU NCY MHz
FREQUENCY MHz
ENTRY DISPLAY
Figure BA-7. YO Diagnostics
[SHIFT] [M2]
(HP-IB: SHM2) is a service diagnostic that shows, from left to right, what the
band number and the YIG oscillator (YO) frequency should be. See Figure 8A-7, "YO
Diagnostics."
PLL #2 Frequency
PLL #3 Frequency
UNLEVELED
START
w
"
STOP
°'
OVERMOD
REMOTE
'""
FAULT
OVEN
EXT REF
UNLK
FREQUE
MHz
FREQU NCY MHz
ENTRY DISPLAY
Figure BA-8.
20-30 Loop Diagnostics
[SHIFT] [M3]
(HP-IB: SHM3) is a service diagnostic that shows, from left to right, what the PLL
#2 VCO frequency and the PLL #3 upconverter frequency should be. Refer to Figure 8A-8,
"20-30 Loop Diagnostics."
[SHIFT] [M4]
(HP-IB: SHM4) activates 32 service diagnostic routines which test several of the
DA Cs and control circuitry on the A27 level control, A28 SYTM driver, A57 marker bandcross,
and A58 sweep generator. This diagnostic also allows the results of the self-test, run at power on
and after on
[INSTR PRESET],
to be displayed in the front panel
ENTRY DISPLAY.
For more
information, refer to the A60 Processor assembly troubleshooting in the Controller Section.
[SHIFT] [M5]
(HP-IB: SHM5) turns off any of the above diagnostic routines and restores the
displays to their previous condition.
[SHIFT] [XTAL]
(HP-IB: SHA2) activates a band crossing diagnostic function. Press
[SHIFT]
[XTAL]
to enable the diagnostic, then press
[SHIFT] [INT],
which will cause the HP 8340A to
sweep to the first band crossing point and stop at that point. Press
[SHIFT] [INT]
again and the
HP 8340A will sweep to the next band crossing point and stop at that point. Continue pressing
[SHIFT] [INT]
to single-step through each of the bands. Press
[SHIFT] [XTAL]
again to cancel
this diagnostic routine.
[SHIFT] [INT]
(HP-IB: SHAl) allows single-stepping through each frequency band, and is used
with
[SHIFT] [XTAL]
for a band crossing diagnostic routine. See
[SHIFT] [XTAL]
for an
explanation of how these two shifted functions interact.
[SHIFT] [RF]
(HP-IB: SHRF) disables the ALC (automatic leveling control) to allow direct
control of the linear modulator circuit. This is useful when very narrow pulses are being
8-39
Scans by HB9HCA and HB9FSX
Содержание 8340A
Страница 1: ...Scans by HB9HCA and HB9FSX ...
Страница 113: ...Scans by HB9HCA and HB9FSX ...
Страница 187: ...Scans by HB9HCA and HB9FSX ...
Страница 198: ...Scans by HB9HCA and HB9FSX ...
Страница 269: ...Scans by HB9HCA and HB9FSX ...
Страница 296: ...Scans by HB9HCA and HB9FSX ...
Страница 320: ...Scans by HB9HCA and HB9FSX ...
Страница 321: ...Scans by HB9HCA and HB9FSX ...
Страница 322: ...Scans by HB9HCA and HB9FSX ...
Страница 323: ...Scans by HB9HCA and HB9FSX ...
Страница 324: ...Scans by HB9HCA and HB9FSX ...
Страница 325: ...Scans by HB9HCA and HB9FSX ...
Страница 326: ...Scans by HB9HCA and HB9FSX ...