Model 8340A - Service
Gain Adjust
F
After being filtered, the
IF
signal is applied to a
common-emitter amplifier Q6 which has adjustable gain.
IF
Gain
adjustment Rl can be adjusted to provide from
5
dB to 20 dB of
gain. This is used to adjust the
IF
signal to the proper level
for comparison to the 20-30 signal in the A49
YO
Phase Detector.
Output Amplifier
G
The
IF
signal is further amplified by output amplifier QS and Ql.
Gain is approximately 21 dB. This provides the proper signal
level to drive the �49
YO
Phase Detector.