8-16
Model 8340A
-
Service
INSTRUMENT PRESET CALIBRATION CONSTANT VERIFICATION SEQUENCE
CALIBRATION CONSTANTS
DATA
CHECKSUM
WORKING (RAM)
2
3
_____
.,..
96
PROTECTED (EE-PROM)
2
3
_____
.,.
9 6 9 7 9 a 9 9
DEFAULT (UY-EPROM)
2
3
_____
.,..
96
NO
WRITE DEFAULT
DATA, CAL. CONST.
!
THROUGH 99, INTO
WORKING AREA
LIGHT 'FAULT'
DISPLAY 'CAL
DEFAULTED"
INSTRUMENT
CONTINUES
OPERATION
(TROUBLESHOOT 'FAULT')
WRITE PROTECTED
DATA, CAL. CONST.
1
THROUGH 99, INTO
WORKING AREA
SUM CAL. CONST.
1
THROUGH 98
THEN COMPLEMENT
NO
YES
INSTRUMENT
PRESET
(NOT POWER UP)
DISPLAY
'CALIBRATION
RESTORED'
CONTINUE
NORMAL
OPERATION
YES
CONTINUE
NORMAL
OPERATION
Figure
BA-3.
Instrument Preset Calibration Constant Verification Sequence