100
Maintenance
7RFRQQHFWWKHYDFXXPJDXJHFRQWUROOHU
To connect the vacuum gauge controller
Materials needed:
Gauge controller (HP 59864B)
Triode gauge cable (8120-6573)
The high-vacuum gauge controller is required for operating the MSD in CI mode.
1
Connect the triode gauge cable to the triode gauge tube.
2
Connect the other end of the triode gauge cable to the gauge controller.
3
Connect the power cord to the gauge controller.
4
Connect the other end of the power cord to an appropriate electrical
outlet.
5
Pump down the MSD.
W A R N I N G
1HYHUFRQQHFWRUGLVFRQQHFWWKHFDEOHIURPWKHWULRGHJDXJHWXEHZKLOHWKH06'LVXQGHU
YDFXXP5LVNRILPSORVLRQDQGLQMXU\GXHWREURNHQJODVVH[LVWV
W A R N I N G
1HYHUFRQQHFWRUGLVFRQQHFWWKHFDEOHIURPWKHWULRGHJDXJHWXEHZKLOHWKHFRQWUROOHULV
WXUQHGRQ'DQJHURXVYROWDJHVDUHSUHVHQWRQWKHFDEOH
C A U T I O N
Do
QRW
use an HP 59864
A
(older model) triode gauge controller with the HP 5973 CI MSD.