Troubleshooting
$LUOHDNV
71
Always
look for small air leaks when setting up methane flow. Run the
methane
pre-tune. The abundance of
m/z
19 (protonated water) should be less than that of
m/z
17. If the MSD was just pumped down, look for the abundance of
m/z
19 to be
decreasing.
There should not be any peak visible at
m/z
32 (O
2
). This almost always indicates
an air leak.
Special negative CI
notes
Since NCI is so extremely sensitive, air leaks that are not detectable in EI or PCI
can cause sensitivity problems in NCI. To check for this kind of air leak in NCI, inject
OFN. The base peak should be at m/z 272. If the abundance of m/z 238 is much
greater than that of m/z 272, you have an air leak.