Troubleshooting
3HDNDWP]
Peak at
m/z
32
A visible peak at
m/z
32 in methane pre-tune often indicates air in the system.
Possible Cause
Residual air from recent venting — check for water indicated by a large peak at
m/z
19.
Action
Bake out the system as described in the
Maintenance
chapter of this manual.
Possible Cause
New or dirty reagent gas supply tubing.
Action
Purge the reagent gas supply lines and flow module
for at least 60 minutes
as
described in the
Maintenance
chapter of this manual.
Possible Cause
Air leak.
Action
Check for leaks and correct any that you find. See the
Leaks
section at the end of
this chapter for more information. After all leaks have been corrected, clean the ion
source.
Possible Cause
Contaminated reagent gas supply. Suspect this if you have recently replaced your
gas tank, and you have ruled out air leaks.
Action
Replace the reagent gas supply.
Possible Cause
The capillary column is broken or disconnected.
Action
Inspect the capillary column. Make sure it is not broken and it is installed correctly.
Possible Cause
Saturated methane / isobutane gas purifier.
Action
Replace the gas purifier.
Содержание 5973 MSD
Страница 1: ...H Hardware Manual HP 5973 MSD PCI NCI ...
Страница 6: ...6 ...
Страница 7: ...1 General Information Specifications 9 Reagent gases and regulators 10 ...
Страница 13: ... Hardware Reference 13 ...
Страница 25: ... Chemical Ionization Theory 3RVLWLYH WKHRU 25 ...
Страница 31: ... Chemical Ionization Theory 1HJDWLYH WKHRU 31 ...
Страница 34: ...34 ...
Страница 92: ...92 ...
Страница 103: ...8 Parts GC MSD interface parts 106 CI ion source parts 108 Flow control module parts 110 Miscellaneous parts 113 ...
Страница 114: ...114 ...
Страница 120: ...120 ...
Страница 122: ...H Manual Part Number G1999 90001 Copyright 1998 Hewlett Packard Printed in USA 2 98 ...